Servicing the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a
week.
See also: Instructions for Use KaVo QUATTROcare PLUS 2124 A
Note
Handpieces must be taken off the service couplings before the chuck
service can be started and performed.
▶
Remove the service coupling chuck from the side hatch of the
QUATTROcare PLUS and attach it to coupling service point four, on
the far right. A MULTIflex adapter must be mounted there.
Reprocessing steps in accordance with ISO 17664