If the passability of the spray tube cannot be determined after the
manual rinsing procedure, the dental product must be replaced.
Brush the spray tube under running tap water water for at least 20 sec‐
onds using a medium-hard toothbrush.
The non-fixing preliminary cleaning is a central constituent and must be
carried out prior to the automatic reconditioning.
In the KaVo QUATTROcare CLEAN, a validated interior cleaning of the
spray tube can be achieved by using the surgical adapter coupling.
Reprocessing steps in accordance with ISO 17664
38