Instructions for use For SONICflex tips paro - 0.571.0371, paro A - 1.006.2020

4 First use | 4.1 Inserting the SONICflex tips

4 First use

 WARNING

Hazard from non-sterile products.

Risk of infection to the care provider and patient.

▶ Before first use and after each use, prepare and sterilise the medical device.

4.1 Inserting the SONICflex tips

▶ Insert the desired tip with the tip end pointing down into the torque wrench and

screw it into the handpiece by turning it clockwise.

 CAUTION

Hazard from a tip which has been incorrectly inserted in the torque wrench.

This may result in a risk of injury for the user.

▶ When inserting the tip into the torque spanner, make sure that the tip's end al‐

ways faces into the opening of the torque spanner.

The torque wrench is for changing the working tips of the SONICflex and to protect
against injury. The torque wrench can be screwed in more quickly by holding it at the
rear, thin grip area ①. The large diameter ② is used for tightening and removal.

Note

The tip is properly gripped when the torque wrench slips.

Note

When the SONICflex is in the holder, the torque wrench should be placed on the tip

to protect against injury.

4.2 Removing the SONICflex tip

▶ Place the torque wrench on the SONICflex tip, and unscrew it anticlockwise.

9 / 18

Summary of Contents for SONICflex tips paro

Page 1: ...Instructions for use For SONICflex tips paro 0 571 0371 paro A 1 006 2020 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...2 Instructions for use 10 6 Preparation methods according to ISO 17664 12 6 1 Preparation at the site of use 12 6 2 Preparation before cleaning 12 6 3 Cleaning 12 6 3 1 Manual cleaning of the exterior...

Page 4: ...afely Symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Can be steam sterilised at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Thermodisinfe...

Page 5: ...indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER...

Page 6: ...ng on them gently with your thumb or forefinger In addition expose the tips to approx 10 N 1 kg mechanical load without func tion CAUTION Hazard from worn SONICflex paro tips SONICflex paro tips can b...

Page 7: ...with KaVo SONICflex LUX 2000 L LX N NX 2000 NM LM 2004 LM 2003 2003 L xx Tip identification by 2 digit number Tips with short thread for use with KaVo SONICflex quick 2008 xxA Tip identification by 2...

Page 8: ...Instructions for use For SONICflex tips paro 0 571 0371 paro A 1 006 2020 3 Product description 3 3 Transportation and storage conditions 8 18 Protect from moisture...

Page 9: ...in the torque wrench This may result in a risk of injury for the user When inserting the tip into the torque spanner make sure that the tip s end al ways faces into the opening of the torque spanner...

Page 10: ...nd premature breakage of the tip Level 3 should only be used briefly 5 2 Instructions for use This tip set can be used for initial treatment and regular recall Regular recall treat ments are very impo...

Page 11: ...y to identify The SONICflex paro tips are designed for use in the following areas 60 60 A SONICflex paro straight no 60 60 A all quadrants 61 61 A SONICflex paro left angled no 61 61 A mandible left v...

Page 12: ...bit bath as the fine capillaries would corrode badly making it impossible to rinse them under running water Note If a sterile coolant is used the SONICflex tips must be rinsed with spray water after...

Page 13: ...ith tip screwed on into the adapter for interior cleaning in the thermodisinfector e g Miele accessories for dental silicon adapters The program settings and cleansers and disinfectants that must be u...

Page 14: ...ONICflex instru ment KaVo recommends thermodisinfectors in accordance with ISO 15883 1 e g Miele G 7781 G 7881 Validation was performed with the program VARIO TD the cleaner neodisher mediclean the ne...

Page 15: ...product Immediately remove the product from the steam steriliser after the sterilisation cy cle Autoclave with a triple prevacuum for at least 4 minutes at 134 o C 1 273o F 1 8 Autoclave using the gr...

Page 16: ...0 571 0371 paro A 1 006 2020 7 Tools 6 8 Storage 16 18 7 Tools Obtainable from the dentalmed specialist supplier Material summary Mat no Torque wrench 1 000 4887 Nozzle needle 0 410 0911 Sterilisatio...