Instructions for use For SONICflex tips paro - 0.571.0371, paro A - 1.006.2020
2 Safety | 2.2 Safety instructions
6 / 18
CAUTION
Hazard from using non-KaVo equipment.
This can damage the product and cause malfunction.
▶ The tips may not be used on non-KaVo equipment.
Note
When the SONICflex is in the holder, the torque wrench should be placed on the tip
to protect against injury.
CAUTION
Breakage of the SONICflex tips.
A fracture may occur as a result of long-term use or damage (dropped to the floor or
mechanical change of the original shape).
▶ Before each use, check the tips by pressing on them gently with your thumb or
forefinger.
▶ In addition, expose the tips to approx. 10 N (1 kg) mechanical load without func‐
tion.
CAUTION
Hazard from worn SONICflex paro tips.
SONICflex paro tips can break or become contaminated.
▶ We recommend changing the SONICflex para tips every 9-12 monthsSONICflex
paro Spitzen auszutauschen.
CAUTION
Risk of injury and infection when changing the SONICflex tips.
This can substantially endanger the user.
▶ Use gloves or finger stalls whenever you check, insert or remove the SONICflex
tips.
CAUTION
The SONICflex tips may break if the wrong or excessive output is used.
This can endanger the user and patient.
▶ Do not use the wrong or excessive output setting.
Note
Regularly check the wear of the tip with the tip test card (Mat. no. 1.001.6958).