Instructions for use For SONICflex tips paro - 0.571.0371, paro A - 1.006.2020
6 Preparation methods according to ISO 17664 | 6.6 Packaging
6.6 Packaging
Note
The sterilisation bag must be large enough for the tip so that the bag is not stretch‐
ed.
The quality and use of the sterilised product packaging must satisfy applicable
standards and be suitable for the sterilisation procedure.
Note
The SONICflex can also be sterilised in the sterilisation tray.
▶ Individually weld the medical device in the sterilised item packaging (such as Ka‐
Vo STERIclave bags Mat. no. 0.411.9912).
▶ Tips with short thread and marked A only fit in green mounts.
Tips with long thread fit in blue and green mounts.
6.7 Sterilisation
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisation cy‐
cle!
Autoclave with a triple prevacuum for at least 4 minutes at
134
o
C ± 1 (273
o
F ± 1.8)
Autoclave using the gravitation method for at least 10 minutes at
134
o
C ± 1 (273
o
F ± 1.8)
Autoclave using the gravitation method for at least 60 minutes at
121
o
C ± 1 (250
o
F ± 1.8)
Follow the manufacturer's instructions for use.
Autoclave with a pre-vacuum for least four minutes at 134°C ± 1°C
(273
o
F ± 1.8
o
F)
Drying time: 20 min.
Autoclave with the gravitation method for at least 10 minutes at 134℃ ±1
o
C ± 1
o
C
(273
o
F ± 1.8
o
F)
Drying time: 30 min.
Autoclave with the gravitation method for at least 60 minutes at 121
o
±1
o
C
(250
o
F ± 1.8
o
F)
Drying time: 30 min.
Follow the manufacturer's instructions for use.
6.8 Storage
Prepared products must be stored, protected from germs (as far as possible) and
dust, in a dry, dark, cool room.
Note
Comply with the expiry date of the sterilised items.
15 / 18