![KaVo SCAN EXAM User Manual Download Page 47](http://html1.mh-extra.com/html/kavo/scan-exam/scan-exam_user-manual_3959068047.webp)
SCAN EXAM
A-1
A. Technical Speci
fi
cations
Appendix A. Technical Speci
fi
cations
A.1 Unit
INTRAORAL IMAGING PLATE UNIT
eXam5
Classification IEC60601-1
- Class 1 or 2 equipment depending on the classification
of the PSU. No applied part
- Continuous operation
- IPX0 (enclosed equipment without protection against
ingress of liquids
Indications for use
The Scan eXam imaging system is indicated for
capturing, digitization and processing of intra oral x-ray
images stored in imaging plate recording media.
Intended use
The Scan eXam system is intended to be used only by
dentist and other qualified dental professionals to process
x-ray images exposed to the imaging plates from the
intraoral complex of the skull.
Laser Safety Classification
CLASS 1 LASER PRODUCT
EN 60825-1 :2007
Dimensions (H x W x D)
196 mm x 215 mm x 382mm (7.7in x 8.5in x 15.0in)
Weight
9.8 – 10.3 kg (21.6 – 22.7 lb)
Power supply unit (PSU)
POWERBOX EXM 80 5121
Operating voltage
24 VDC (PSU: 100 – 240 VAC, 50/60 Hz)
Operating current
Less than 1.5 A
Pixel size, selectable
30
μ
m (Super), 60
μ
m (High)
Bit depth
14 bits grayscale
Theoretical resolution
16.7 lp/mm
Interface cable
For Ethernet connections, use an unshielded CAT6
LAN cable, so that multiple chassis must not be
connected! The PC / Ethernet switch to which unit is
connected to should be approved appropriately (e.g.
EN 60950, IEC 60950, UL 60950).
Operating environment
+10
°
C - +40
°
C, 30 – 90 RH%, 700 – 1060 mbar
Storage / transportation
environment
-10
°
C – +50
°
C, 0 – 90 RH%, 500 – 1080 mbar