Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550
6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting
18 / 28
4
Take lamp from the mount.
4
Insert the new lamp very carefully.
6.2.2 Replacing the O-rings
CAUTION
Missing or damaged O-rings.
Malfunction and premature failure.
4
Make sure that all O-rings are on the coupling and are undamaged.
NOTE
Stop working if the O-ring is missing or damaged.
CAUTION
Improper care of the O-rings.
Malfunction or complete failure.
4
Do not use Vaseline or other grease or oil.
NOTE
The O-rings on the coupling may only be lubricated with a cotton ball wetted
with KaVo Spray.
NOTE
If the connection to the instrument is not leak-proof, replace all O-rings.
4
Press the O-ring between your fingers to form a loop.
4
Push the O-ring to the front, and remove it.
4
Insert new O-rings into the recesses and spray them with KaVo Spray.