background image

Air consumption

up to 60 Nl/min

Idle torque

2.3 Ncm at 2.2 bar (32 psi)

The INTRAmatic L motor can be attached to all hoses with Borden 2-hole connection.

With INTRAmatic 1 coupling.

All straight and contra-angle handpieces with INTRAmatic connection ISO 3964 can be attached.

 

 

 

Product description

23

33.3

   

Transportation and storage conditions

CAUTION

It is hazardous to start up the medical device after it has been stored strongly refrigerated.

This can cause the medical device to malfunction.

▶ Prior to start-up, very cold products must be heated to a temperature of 20°C to 25°C

(68°F to 77°F).

 

Product description

24

Temperature: -20°C to +70°C (-4°F to +158°F)

Relative humidity: 5% RH to 95% RH absence of condensation

Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)

Protect from moisture

 

Product description

25

33.4

   

Operating conditions

Temperature: +5 °C to +40 °C (41 °F to 104 °F)

Maximal rel. humidity: < 80 %

Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)

 

 

Product description

26

Summary of Contents for L-MOTOR 181 DB INTRAmatic I

Page 1: ...y instructions 10 2 1 1 Description of safety instructions Warning symbol 10 2 1 2 Description of safety instructions Structure 11 2 1 3 Description of safety instructions Description of danger levels...

Page 2: ...leaning Automated external cleaning 44 7 2 3 Cleaning Manual internal cleaning 45 7 2 4 Cleaning Automated internal cleaning 45 7 3 Disinfection 46 7 3 1 Disinfection Manual external disinfection 47 T...

Page 3: ...t information for users and service technicians Action request User instructions 8 CE mark European Community A product with this mark meets the requirements of the relevant EC directives i e the stan...

Page 4: ...three levels of danger will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moderate injuries Safety 12 WARNING WARN...

Page 5: ...ording to instructions before long periods of nonuse The following individuals are authorized to repair and service KaVo products Technicians at KaVo branches throughout the world Technicians speciall...

Page 6: ...ing to these regulations this medical device may only be used for the described application by a knowledgeable user The following must be observed the applicable health and safety regulations the appl...

Page 7: ...dical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F t...

Page 8: ...sed air according to ISO 7494 2 First use 28 Insert the test manometer M Mat no 0 411 8531 between hose and motor and check the fol lowing pressures Drive pressure 2 2 to 2 7 bar 32 to 39 psi recommen...

Page 9: ...st the water amount for spray cooling with the spray ring fully opened to approx 80 cm3 min Comply with the settings recommended by the manufacturers Operation 32 Rotate the spray tap of the hose or d...

Page 10: ...Rotate the regulating sleeve to regulate between idle in clockwise rotation or counter clockwise rotation and maximal speed in a continuous manner Operation 36 6 6 Troubleshooting 6 1 Check for malfun...

Page 11: ...rings forwards and remove them Troubleshooting 40 Insert new O rings into the grooves and spray them with KaVo spray Press the O rings between your fingers to form a loop Slide the O rings forwards an...

Page 12: ...Wipe the surface of the motor with a water soaked cotton cloth and then dry with a soft cot ton cloth 7 2 2 Cleaning Automated external cleaning Not applicable Preparation methods according to ISO 17...

Page 13: ...the disinfectant manufacturer Follow the instructions for use of the disinfectant 7 3 2 Disinfection Manual internal disinfection Not applicable Preparation methods according to ISO 17664 48 7 7 3 3...

Page 14: ...ter each cleaning Unscrew the motor from the hose Cover the product with the Cleanpac bag Use an appropriate cannula and apply spray into the larger of the two tubes for one second Preparation methods...

Page 15: ...7 Storage Reprocessed products should be stored protected from dust with minimum exposure to germs in a dry dark and cool space Preparation methods according to ISO 17664 56 8 8 Auxiliary equipment A...

Page 16: ...wear improper handling cleaning or maintenance non compliance with operating main tenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electri...

Page 17: ...0 589 2152 kb 20111020 03 en...

Reviews: