Instructions for use INTRAflex LUX 2 2313 LN - 0.535.6220

2 Safety | 2.1 Infection hazard

6 / 26

2 Safety

The instructions for use are a component of the product and must be read care-
fully prior to use and be accessible at all times.
The device may only be used in accordance with the intended use, any other
type of use is not permitted.

2.1 Infection hazard

Patients, users or third parties could be infected by contaminated medical
devices.

Take suitable personal protective measures.

Follow the instructions for use of the components.

Before the initial startup and after each use, reprocess and sterilise the
medical device and accessories accordingly.

Carry out the cleaning and sterilisation as described in the instructions for
use. The procedure has been validated by the manufacturer.

It is essential to ensure the effectiveness of the cleaning and sterilisation in
the case of deviation in procedure.

Prior to disposal, the product and accessories must be appropriately repro-
cessed or sterilised.

If there is any injury to soft tissue, do not continue treatment in the oral
cavity with compressed air-driven instruments.

2.2 Technical condition

A damaged device or components could injure patients, users and third parties.

Only operate devices or components if they are undamaged on the outside.

Check that the device is working properly and is in satisfactory condition
before each use.

Have parts with sites of breakage or surface changes checked by the Ser-
vice.

If the following defects occur, stop working and have the service personnel
carry out repair work:

▪ Malfunctions

▪ Damage

▪ Irregular running noise

▪ Excessive vibration

▪ Overheating

▪ Dental bur or diamonds is not firmly locked in the handpiece

Observe the following instructions in order to guarantee optimum functioning
and prevent material damage:

Service the medical device with care products and systems regularly as de-
scribed in the instructions for use.

The device should be cleaned, serviced and stored in a dry location, accord-
ing to instructions, if it will not be used for a longer period.

Summary of Contents for INTRAflex LUX 2 2313 LN

Page 1: ...Instructions for use INTRAflex LUX 2 2313 LN 0 535 6220...

Page 2: ...stributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo c...

Page 3: ...4 Remove handpiece or contra angle drill from handpiece 2313 LN 14 5 5 Modification of the handpiece 2313 LN for contra angle drills 14 6 Troubleshooting 16 6 1 Replacing the O rings on the coupling...

Page 4: ...For scheduling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll free Direct Customer Service 1 888 ASK KAV...

Page 5: ...and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material...

Page 6: ...oduct and accessories must be appropriately repro cessed or sterilised If there is any injury to soft tissue do not continue treatment in the oral cavity with compressed air driven instruments 2 2 Tec...

Page 7: ...piece head or instrument cover Do not use the medical device as a light probe Use an appropriate light probe for illumination of the oral cavity or site of preparation After treatment place the medica...

Page 8: ...sale by or on the order of a health care professional dentist For dental use only Proper use According to these regulations this product may only be used for the described application by a properly t...

Page 9: ...s drills can be used Attachable to all MULTIflex LUX MULTIflex LED couplings 3 3 Transportation and storage conditions Do not store in a refrigerated environment Temperature 20 C to 70 C 4 F to 158 F...

Page 10: ...ake sure that the supply of cooling air is dry clean and uncontaminated according to ISO 7494 2 4 1 Mount the MULTIflex LUX MULTIflex LED coupling WARNING Detachment of the medical device during treat...

Page 11: ...Instructions for use INTRAflex LUX 2 2313 LN 0 535 6220 4 Start up and shut down 4 2 Checking the amount of water 11 26...

Page 12: ...upling and the medical device and check the following pressures Drive pressure drive pressure recommended return air pressure spray water pressure and spray air pressure Drive pressure 2 5 to 3 0 bar...

Page 13: ...pling can destroy the high pressure lamp or the LED of the coupling or reduce its service life Make sure that the turbine is accurately coupled and firmly seated on the coupling Mount the medical devi...

Page 14: ...insert to prevent injury and infection Rotate the clamping ring sleeve in the direction of the arrow to the stop and insert the handpiece cutter or polisher in the ring Turn the clamping ring back int...

Page 15: ...Instructions for use INTRAflex LUX 2 2313 LN 0 535 6220 5 Operation 5 5 Modification of the handpiece 2313 LN for contra angle drills Use the enclosed hook to remove the bur stop 15 26...

Page 16: ...th a cotton ball wetted with KaVo Spray Press the O ring between your fingers to form a loop Push the O ring to the front and remove it Insert new O rings into the grooves 6 2 Cleaning the spray nozzl...

Page 17: ...ice as soon as possible after treatment To minimise the risk of infection during reprocessing always wear protect ive gloves Remove the tool from the medical device Remove all residual cement composit...

Page 18: ...ility of the materi als The microbiological efficacy must be ensured by the disinfectant manufac turer and proven by an expert opinion CaviCide made by Metrex Consumables required Cloths for wiping th...

Page 19: ...Only possible for the INTRAflex contra angle KaVo recommends servicing the chucking system once weekly Remove the tool from the medical device Position the tip of the spray nipple in the opening and...

Page 20: ...rvice symbol Note Close the chuck service mode Option 1 Place the dental handpieces in the QUATTROcare PLUS 2124 A close the front door and start theservice procedure Option 2 After three minutes with...

Page 21: ...with pre vacuum at least 3 minutes at 135 C 275 F Drying time 16 min 2 Steriliser using the gravity method at least 10 minutes at 135 C 275 F Drying time 30 min Remove contra angle handpieces and tur...

Page 22: ...mary Mat No Instrument stand 2151 3 005 5204 Insert for INTRAflex 3 005 3089 Cellulose pad 100 units 0 411 9862 Cleanpac 10 units 0 411 9691 Nozzle pin 0 410 0931 Bur stop 0 524 0892 Hook 0 410 1633 M...

Page 23: ...their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or corrosi...

Page 24: ......

Page 25: ......

Page 26: ...1 009 4692 Fk 20190214 09 en...

Reviews: