Instructions for use INTRAflex LUX 2 2313 LN - 0.535.6220
7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging
20 / 26
Note
Handpieces must be taken off the service couplings before the chuck service
can be started and performed.
▶
Close the front door and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecut-
ively.
ð
The device is in chuck service mode.
▶
Remove the service coupling of the chuck from the side hatch of the QUAT-
TROcare PLUS and attach it to coupling service point four, on the far right.
A MULTIflex adaptor must be mounted there.
▶
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
▶
Press the button marked with the chuck service symbol.
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start theservice procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
See also:
2
Servicing with KaVo QUATTROcare PLUS
7.5 Packaging
Note
The sterilisation bag must be large enough for the instrument so that the bag
is not stretched.
The quality and use of the sterilisation packaging must comply with applic-
able standards and be suitable for the sterilisation procedure!
▶
The medical device must be packed before sterilisation.
7.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with ISO 17665-1
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.
NOTICE
Contact corrosion due to moisture.
Damage to product.
▶
Immediately remove the product from the steam steriliser after the steril-
isation cycle.