Instructions for use INTRAflex LUX 2 Programm 2307 LN - 0.535.6190, 2320 LN - 0.535.6180, 2325 LN -
0.535.6210
7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging
KaVo recommends servicing the chuck system once a week using the chuck
servicing program integrated in the device.
Note
Handpieces must be taken off the service couplings before the chuck service
can be started and performed.
▶
Close the front door and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecut-
ively.
ð
The device is in chuck service mode.
▶
Remove the service coupling of the chuck from the side hatch of the QUAT-
TROcare PLUS and attach it to coupling service point four, on the far right.
A MULTIflex adaptor must be mounted there.
▶
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
▶
Press the button marked with the chuck service symbol.
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start theservice procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
See also:
2
Servicing with KaVo QUATTROcare PLUS
7.5 Packaging
Note
The sterile goods package must be large enough for the product so that the
packaging is not stretched.
The quality and use of the packaging of the items to be sterilised must satisfy
the applicable standards and be appropriate for the sterilisation process!
▶
The medical device must be packed before sterilisation.
7.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with ISO 17665-1
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.
23 / 28