7.2.4
Cleaning: Automated internal cleaning
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
e.g.. Miele G 7781 / G 7881.
(Validation was performed with the program "VARIO-TD", the detergent "neodisher
®
mediclean", the
neutraliser "neodisher
®
Z" and rinse "neodisher
®
mielclear" and only refers to the material
compatibility with KaVo products.)
▶
The program settings and cleansers and disinfectants that must be used can be found in the
instructions for use of the thermodisinfector.
Preparation methods according to ISO 17664
52