6.4.2 Cleaning: Automated external cleaning
KaVo recommends thermodisinfectors in accordance with ISO 15883-1
e.g. Miele G 7781/G 7881.
6.4.3 Cleaning: Manual cleaning of the inside
For a medical device with KaVo CLEANspray and KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
▶
Then rinse for 3-5 seconds with KaVo DRYspray.
See also: Instructions for use for the KaVo CLEANspray/KaVo DRYspray.
Preparation methods according to ISO 17664
38
Summary of Contents for INTRA 3610 N3
Page 1: ...Instructions for use INTRA surgical handpiece 3610 N3 REF 0 524 5620...
Page 21: ...Air pressure 700 hPa 10 psi to 1060 hPa 15 psi Protect from moisture Product description 21...