The medical device must be dry when transporting it to be prepared.
(Do not place it in a solution or the like).
▶
Prepare the medical device directly before treatment.
▶
Remove cutters or burs from the medical device.
6.2 Disassemble the medical device
▶
Remove cutters or burs from the medical device.
▶
Grip the sleeve ② and remove the grip sleeve ① by pulling it forward.
Preparation methods according to ISO 17664
36