Technician's instructions GENTLEpower LUX Contra-angle 25 LP - REF 1.001.2849
4 Testing | 4.1 Testing the GENTLEpower LUX Angle Piece 25 LP
16 / 20
4 Testing
4.1 Testing the GENTLEpower LUX Angle Piece 25 LP
4.1.1 Test set-up
① Test device 2182 KL (Mat. no.
0.411.8540)
② Leakage air hose
③ Motor hose
④ Pressure gauge (Mat. no. 1.003.1050)
⑤ Motor
4.1.2 Test instruction
Note
The following evidence must be documented:
- Customer's address and repair dates
- Material number or sales number of the product
- Serial number of the product
- Spare parts material numbers
- Name, date, signature or personnel number of repair/test engineer
Requirement
- Amount of cooling air: 28 ± 4 Nl/min
- Drive speed: max. 40,000 rpm
The pressures are set on the test device.
Test equipment
Name
Test drill (Mat. no. 0.410.1702)