Technician's instructions GENTLEpower LUX Contra-angle 25 LP - REF 1.001.2849
3 Assembly | 3.1 Assembly of the GENTLEpower LUX Angle Piece 25 LP
Tool
Description
Loctite 306 adhesive (Mat. no. 0.359.2044
Note
Do not glue spray pipes when gluing the fibre-optic conductor.
▶ Fit a new O-ring ① to the top of spray pipe ② and a new O-ring ① to the top of
spray pipe ③.
▶ Fit a new O-ring ⑦ to the bottom of spray pipe ② and a new O-ring ⑦ to the bot‐
tom of spray pipe ③.
▶ Push the head casing ⑤ into the conical sleeve ⑧ to insert spray pipe ② and
spray pipe ③ into the head casing ⑤.
▶ Insert the fibre-optic bundle ④ into the conical sleeve ⑧ and slide over spray pipe
②, spray pipe ③, and head casing⑤.
▶ Apply a layer of Loctite 648 to the glued joints ⑨ on the elbow sleeve.
▶ Attach the conical sleeve ⑧ to the elbow sleeve ⑩ and wipe off any excess adhe‐
sive.
▶ Carefully push the axle ⑫, bearing cartridge ⑪ and insert ⑬ all the way into the
elbow sleeve ⑩.
▶ Screw the threaded ring ⑭ into the elbow sleeve ⑩ and tighten to 180 Ncm with
the wrench, the universal bit holder and the torque handle.