7 Technical data
38
FOCUS
Table 3: RF immunity of non-life-support equipment or system IEC 60601-1-2 ed.4
The device is suitable for use in the specified electromagnetic environment. The purchaser or user
of the device should assure that it is used in an electromagnetic environment as described below:
Immunity Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic Environment
Conducted RF
IEC 61000-4-6
3V 150kHz to 80
MHz, 6 V in ISM
and amateur radio
bands between
0,15 MHz and 80
MHz according to
table 60601-1-2.
Ed4. Table 5/8.
3V 150kHz to 80
MHz, 6 V in ISM
and amateur radio
bands between
0,15 MHz and 80
MHz according to
table 60601-1-2.
Ed4. Table 5/8.
Portable and mobile RF
communications equipment are used no
closer to any part of the device including
cables, than the recommended
separation distance calculated from the
equation appropriate for the frequency
of the transmitter.
Recommended Separation Distance:
150kHz-80MHz
80MHz to 800 MHz
800 MHz to 2,7 GHz
Where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation distance in
meters (m). Field strengths from fixed
RF transmitters, as determined by an
electromagnetic site survey,* are less
than the compliance level in each
frequency range.**
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF IEC
61000-4-3
10 V/m 80 MHz to
2,7 GHz Immunity
to proximity fields
from RF wireless
communication
equipment, levels
according to table
60601-1-2 ed4
table 9.
10 V/m 80 MHz to
2,7 GHz Immunity
to proximity fields
from RF wireless
communication
equipment, levels
according to table
60601-1-2 ed4 table
9.