KaVo EXPERTmatic LUX E25 L Instructions For Use Manual Download Page 9

Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551

2 Safety | 2.6 Service and repair

9 / 28

2.6 Service and repair

Repairs, servicing and safety checks may only be performed by trained service
personnel. The following persons are authorised to do this:

▪ Service technicians of KaVo branches after the appropriate product training

▪ Service technicians of KaVo authorised dealers after the appropriate product

training

Observe all the following items during servicing work:

4

Have the service and testing tasks carried out according to the Medical
Device Operator Ordinance.

4

After servicing, interventions on and repairs of the device and before re-use,
have the service personnel perform safety checks on the device.

4

Following expiry of the warranty, have the tool holding system checked
once a year.

4

KaVo recommends specifying in-house service intervals where the medical
device is brought to a professional shop for cleaning, servicing and a func-
tion check. Define the service interval depending on the frequency of use.

As a result of the use of NON-KaVo original spare parts during the repair, parts
such as covers may become undone and injure the patient, user or other
people. This may result in aspiration, swallowing of parts and possibly even a
risk of suffocation.

4

Only use spare parts that comply with the specification for repair; original
KaVo spare parts comply with the specification.

NOTE

If a repair is done with NON-KaVo original spare parts, this may constitute a
product modification that leads to the loss of CE conformity. In the event of
damage, the responsibility is with the service company or the operator.

The introduction of a modified product in the market, in which the reasonable
suspicion exists to endanger the safety and health of patients or users, is pro-
hibited by Medical Device Law §4, Abs.1 No. 1 and therefore requires its own
conformity check.

Summary of Contents for EXPERTmatic LUX E25 L

Page 1: ...Instructions for use EXPERTmatic LUX E25 L 1 007 5550 EXPERTmatic E25 C 1 007 5551...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...6 6 Checking for malfunctions and troubleshooting 17 6 1 Checking for malfunctions 17 6 2 Troubleshooting 17 6 2 1 Replacing the O rings on the motor coupling 17 6 2 2 Changing the water filter 18 7 R...

Page 4: ...ng www kavobox com KaVo Technical Service If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions...

Page 5: ...ure range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Original language German Hazard levels Th...

Page 6: ...Instructions for use EXPERTmatic LUX E25 L 1 007 5550 EXPERTmatic E25 C 1 007 5551 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to material damage...

Page 7: ...ely before disposal 4 In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 4 Use gloves or a finger guard when you test insert an...

Page 8: ...any damaged motors 4 Do not use motors for unauthorised purposes 2 4 Accessories and combination with other equipment Use of non authorised accessories or non authorised modifications of the device co...

Page 9: ...ing in house service intervals where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As...

Page 10: ...for the following applications Removal of carious material Removal of fillings Processing of tooth and restoration surfaces Tooth cavity and crown preparations A medical device according to relevant n...

Page 11: ...ling air flow 5 5 to 9 5 Nl min Pushbutton chuck 1 6 mm Insert Dental burs and diamond grinders in accordance with DIN EN ISO 1797 type 3 Can be attached to All INTRA LUX motors and motors with a conn...

Page 12: ...g waste management recycling companies Comply with the comprehensive return system KaVo has had its packaging licensed for this purpose Please comply with the regional public waste disposal system CAU...

Page 13: ...Instructions for use EXPERTmatic LUX E25 L 1 007 5550 EXPERTmatic E25 C 1 007 5551 4 Startup and shut down 4 1 Checking the amount of water 13 28...

Page 14: ...smission handpieces being at tached and if there is a risk of contamination from reflux or back suction the system may also need to be rinsed for 20 to 30 seconds after each patient 5 1 Attaching the...

Page 15: ...ith DIN EN ISO 1797 type 3 are made of steel or hard metal and meet the following criteria Shaft diameter 1 59 to 1 60 mm 0 0626 in to 0 0629 in Overall length max 25 mm Shaft clamping length at least...

Page 16: ...of the dental bur Material damage to dental bur shaft and chuck system reduction of the service life of dental bur and chuck system 4 Do not operate the dental bur at a higher speed than recommended b...

Page 17: ...ctions for use of motor 6 2 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts Parts such as the cover can come loose and cause injury Aspiration swallowing of parts and danger of...

Page 18: ...fection hazard to the dentist and patient 4 Prior to initial startup and after each use reprocess the product and ac cessories CAUTION Overheating of the tooth due to insufficient amount of cooling wa...

Page 19: ...ves 4 Remove the dental bur from the medical device 4 Remove all residual cement composite or blood immediately 4 Do not place in solutions or similar substances 7 2 Manual Reprocessing WARNING Sharp...

Page 20: ...ride containing products Malfunction and material damage 4 Reprocess in a washer disinfector or by hand only KaVo recommends the following products based on the compatibility of the ma terials The mic...

Page 21: ...7 2 5 Manual drying Use KaVo DRYspray for subsequent drying of the air water and gear unit ducts 4 Cover the medical device with the KaVo Cleanpac bag and place it on the corresponding care adapter 4...

Page 22: ...tor 4 For adaptations unless otherwise specified refer to the Instructions for Use of the washer disinfector 7 3 2 Automated drying The drying procedure is usually part of the cleaning programme of th...

Page 23: ...e interior in accordance with German Robert Koch Institute requirements KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before...

Page 24: ...n a sterile pack 7 6 Sterilisation Sterilisation in a steam steriliser autoclave in accordance with EN 13060 EN ISO 17665 1 CAUTION Improper service and care Risk of injury 4 Service regularly with su...

Page 25: ...C 1 007 5551 7 Reprocessing steps in accordance with ISO 17664 7 7 Storage 25 28 7 7 Storage Reprocessed products must be stored appropriately protected from dust in a dry dark and cool space with a l...

Page 26: ...005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0921 Replacement filter 1 002 0271 Wrench for water filter 1 002 0321 O ring 8 3x0 68 0 200 6120 KaVo CLEA...

Page 27: ...aVo shall not be liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintena...

Page 28: ...1 008 1659 bd 20221209 05 en...

Reviews: