Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing
23 / 28
7.4.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation, but no
later than after 30 minutes of operation.
4
Remove the dental bur from the medical device.
4
Cover the medical device with the KaVo Cleanpac bag, and place it on the
corresponding care adapter.
4
Press the spray key for 1 to 2 seconds.
Servicing the chuck
KaVo recommends servicing the chucking system once weekly.
4
Remove the dental bur from the medical device.
4
Position the tip of the spray nipple in the opening, and apply the spray.
4
Press the spray key for 1 to 2 seconds.
7.4.2 Servicing with KaVo QUATTROcare PLUS
Cleaning and servicing device with expansion pressure for internal cleaning of
inorganic residues and optimum care.
(no validated cleaning of the interior in accordance with German Robert Koch
Institute requirements)
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation, but no
later than after 30 minutes of operation.
4
Remove the dental bur from the medical device.
4
Service the device in the QUATTROcare PLUS.
Also refer to:
Instructions for use KaVo QUATTROcare PLUS
Servicing the chuck
KaVo recommends servicing the chuck system once weekly using the chuck
servicing programme integrated in the device.
NOTE
Handpieces must be taken off the service couplings before the chuck service
can be started and performed.
4
Close the front flap and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecut-
ively.
ð
The device is in chuck service mode.
4
Remove the service coupling of the chuck from the side hatch of the
QUATTROcare PLUS and attach it to coupling service point four, on the far
right. A MULTIflex adaptor must be mounted there.
4
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
4
Press the button marked with the chuck service symbol.