Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing
22 / 28
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
7.3.1 Automated internal and external cleaning and
internal and external disinfection
KaVo recommends washer disinfectors in accordance with EN ISO 15883-1,
which are operated using alkaline cleaning agents.
The validations were performed using a Miele washer disinfector and the
"VARIO-TD" programme and a mild alkaline cleaner made by Dr. Weigert.
In addition, KaVo recommends the use of a rinsing agent.
4
For programme settings as well as cleansers and disinfectants, please refer
to the Instructions for Use of the washer disinfector.
4
For adaptations, unless otherwise specified, refer to the Instructions for Use
of the washer disinfector.
7.3.2 Automated drying
The drying procedure is usually part of the cleaning programme of the washer
disinfector.
NOTE
Please comply with the instructions for use of the washer disinfector.
4
In order not to affect the KaVo medical device, make sure that the product
is dry on the inside and outside after completion of the cycle.
4
Remove any residual liquids with KaVoDRYspray.
Also refer to:
4
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
7.4 Care products and systems - Servicing
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
CAUTION
Improper service and care.
Risk of injury.
4
Service regularly with suitable agents.
NOTE
KaVo guarantees the proper function of KaVo products only if the care
products listed by KaVo as accessories are used, since these were tested for
proper use on our products.