Short instructions for use 11 Charging the wireless foot control
35
11 Charging the wireless foot control
Rear of the wireless foot control
Item
No.
Labelling
Function
①
LED display
Status display / charge status display
②
On/Off switch
On/off switch to prevent deep discharge dur‐
ing long periods of non-use. The wireless
foot control can remain switched on at all
times as a matter of principle. The device
must be switched off for transport. The bat‐
tery can also be charged when it is switched
off.
③
Charge socket
Charge socket for the provided charger (Mat.
no. 1.005.4229).
CAUTION
Critical battery level
If the battery reaches a critical charge status, a signal is sounded every time a func‐
tion key is pressed.
▶ Always charge batteries when necessary.
▶ To ensure that the battery of the wireless foot control is always charged in a time‐
ly manner, note the visual and acoustic signals of the wireless foot control when
starting the treatment unit.
Note
Charge the wireless foot control with the charger supplied by KaVo only.
▶ Connect the charger to the wireless foot control.
Residual capacity
Foot control status
Status display / charge
status display
Beep
< 100 %
Idle state
Foot control is on
Flashes green
(approx. 2 second inter‐
vals)
-
Active actuation
Flashes green
(approx. 200 millisecond
intervals)
-