Care instructions ESTETICA E50 Life
3 Replacing and checking cleaning and germ reduction agents | 3.2 Check the OXYGENAL 6 concentration
52 / 58
Note
Substances remaining in the dosing head are not contamination, but rather
agents that are deposited there during extended periods of non-use after
condensation of the water. These deposits redissolve after several dosing
processes.
▶
Screw the water bottle tightly to the treatment centre in a clockwise rota-
tion.
ð
Apply compressed air to the water bottle.
3.2 Check the OXYGENAL 6 concentration
▶
Test the standard concentration 0.02% H
2
O
2
(200ppm): Fill a measuring
cup (40 ml) with 39 ml water and 1 ml treated water.
Summary of Contents for ESTETICA E50 Life