Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
28 / 58
▶
Place the cleaned and disinfected handpieces separately in sterilisation
packages and weld them to be sealed, (e.g. KaVo STERIclave bags
Mat.
no. 0.411.9912
).
135°C
Steriliser with a triple pre-vacuum for at least 4 minutes at 134 °C ± 1 °C
(273.2 °F ± 33.8 °F)
For range of applications, please refer to the manufacturer's Instructions for
Use.
Only for handpieces with fibre optic conductor sleeve
▶
If the fibre optic conductor sleeve
Mat. no. 1.007.4021
looses its bright-
ness due to sterilisation, replace the fibre optic conductor sleeve.
The light source in the handpiece cannot be replaced.
Sterilisation of tips, endo files, file holders, endo
wrenches, torque wrenches
Note
The maximal number of sterilisation cycles must not be exceeded.
Note
The use of hot-air sterilisation and radio-sterilisation is not permissible
(causes destruction of the products). KaVo shall not be held responsible if
non-permissible procedures such as ethylene oxide, formaldehyde, and low
temperature plasma sterilisation are used.
Note
Only cleaned and disinfected products may be sterilised.
Note
Please comply with the current local regulations governing the re-use and
disposal of equipment.