Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
24 / 58
Cleaning of tips, endo files, file holders, endo wrenches,
torque wrenches
Note
For exposure times and concentrations of disinfection agents, please refer to
the instructions of the manufacturers.
▶
Place the products in the disinfection solution at least for as long as spe-
cified by the manufacturer of the disinfection agent.
▶
Remove all contamination from the outer surface by careful brushing using
a soft brush or soft cloth.
▶
Rinse the inside of products thoroughly at least five times using fresh dis-
tilled or deionised water (at least 50 ml).
▶
If the final rinse is not clear or if the product continues to contain visible
contamination, the cleaning process must be repeated.
Disinfection
NOTICE
Malfunctioning from using a disinfectant bath or chloride-containing
disinfectants.
Defects in the device.
▶
Disinfect manually only!
Disinfection of handpieces
Note
For times and concentrations, please refer to the instructions of the manufac-
turers of the cleaning/disinfection agent.
Manual external disinfection
KaVo recommends the following products based on compatibility of the materi-
als. The microbiological efficacy must be confirmed by the disinfectant manu-
facturer.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
▪ Microcide AF Liquid (Schülke & Mayr)
▪ FD 322 (Dürr Dental)
▪ Incidin Liquid (Ecolab)
▶
Wipe the surface with approved disinfectant for disinfection.
▶
Dry the products with filtered compressed air (max. 3 bar).
▶
If required, dry again at a clean place.
▶
Package products immediately after drying (see section on Packaging and
Sterilisation).