Press "Preselect level" button for 4 seconds.
ð
The display switches to the settings menu of the PiezoLED.
▶
Press the "Save" key to save the values. You can save after setting each
value, or after setting all values.
ð
Saving is acknowledged with a tone.
Setting the intensity
▶
Press the button for "Decrease value" to decrease the intensity.
or
▶
Press the button for "Increase value" to increase the intensity.
ð
The intensity is shown on the display.
Define operating mode (PiezoLED only)
Note
The selection of the mode depends on the treatment method and the tip
used. For information about the selection of an operating mode, please refer
to the "Operating modes P1 / P2 / P3 and E" section of the "PiezoLED In-
structions for Use".
▶
Press the "Mode" key to select the operating mode.
Modes P1 / P2 / P3 / E are available for selection.
Setting the cooling level
▶
Adjust the cooling with the selection button for "Cooling status".
Symbol
Function
No cooling
Spray water cooling
Dosing the amount of spray water
CAUTION
Lack of working tip cooling.
Heat damage to tooth or handpiece.
▶
Never work under dry conditions, except in case of tips designed for these
conditions.
▶
Set a minimum flow rate of 6 ml/min. For this purpose, adjust the flow rate
such the water is just between dripping and flowing during irrigation.