Instructions for use ESTETICA E50 Life
3 Product description | 3.11 Technical data
Operating environment
WARNING
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
▶
It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
Floor quality
The quality of the flooring must meet
the load bearing ability for buildings
DIN 1055 page 3 and have a pres-
sure resistance in accordance with
DIN 18560 T 1.
Ambient temperature
+10 to +40
o
C
Relative Humidity
30% to 75% non-condensing
Air pressure
700 hPa - 1,060 hPa
Max. elevation for operation
up to 3,000 m
Maximum loads
Max. patient weight load on patient
chair Standard
185 kg
Max. patient weight load
COMPACTchair
135 kg
Tray holder of the dentist element -
loadable up to
2 kg
Assistant element standard tray -
loadable up to
1 kg
Dentist element - loadable up to
2 kg
Transportation and storage conditions
Ambient temperature
-20 to +55
o
C
Relative Humidity
5% to 95% non-condensing
Air pressure
700 hPa to 1,060 hPa
45 / 162