Instructions for use ESTETICA E50 Life
3 Product description | 3.11 Technical data
Free drainage according to DIN EN
1717 - DVGW certified
Water block DVGW, water bottle
DVGW, register no.: AS-0630BT0111
Water quality
Tap water, cold water connection
Water hardness
1.5 to 2.14 mmol/l
≙
8.4 to 12 °dH
pH
7.2 to 7.8
Customer-site water filtering
80 microns
Water connection
Shut-off valve with brass cone com-
pression screw connection 3/8" to Ø
10 mm provided
Above-floor water connection
min. 50 mm, max. 105 mm with
valve opened
Water inlet pressure
2.0 to 6.0 bar
Water inflow
4 l/min
Diameter of the drain connection
40 mm
Above-floor drain connection
20 mm
Outflow quantity
max. 4 l/min
Slope of water drain pipe
downstream from device: at least 10
mm per metre
Air supply
WARNING
Non-compliance with national guidelines concerning the quality of the
dental air.
Infection hazard.
▶
Note and comply with the national guidelines concerning the quality of the
dental air - if any.
▶
Blow through the air line prior to commissioning.
Air inlet pressure
5.2 to 7 bar
Minimum air flow rate
min. 80 Nl/min
Pressure dew point
< -30 ºC (compressor with
dry air system)
Oil content
< 0.1 mg/m
3
(oil-free compressor)
Contamination
< 100 particles/m
3
at particle sizes
of 1 to 5 µm
Customer-site air filtration
50 microns
Air connection
Shut-off valve with brass cone com-
pression screw connection 3/8" to Ø
10 mm provided
Air connection above floor level
min. 50 mm, max. 105 mm with
valve opened
43 / 162
Summary of Contents for ESTETICA E50 Life
Page 1: ...Instructions for use ESTETICA E50 Life...
Page 161: ......