Key
Settings
S2
Previous message
S3
Next message
S4
Switch to stand-by menu
Error messages in the status display
Malfunction
Cause
Remedy
Display shows: ID 64
Water is shut off.
▶ Turn water on.
Water works leaks strongly.
Water works malfunction
▶ Call a technician.
Display shows: ID 65
Safety switch of bowl suction
has been reached.
▶ Turn external suction on.
▶ Check and clean, if required,
the bowl valve.
Display shows: ID 68
Call for service
▶ Have a service performed.
▶ Call a technician.
Display shows: ID XX
This error is not described in this
chapter.
▶ Call a technician.
Display shows: CAN fail
Internal communication error.
▶ Turn the unit off and on again,
consult a service technician
according to need.