KaVo ESTETICA E30 Short Instructions For Use Download Page 17

The following settings can be changed with the foot control:

▪ Direction of motor rotation

▪ Speed

▪ Cold light intensity

▪ Spray status

▪ Light on/off

6.4

   

KL 703 LED in ENDO mode (optional accessories)

Note
The functions of the ENDO menu are available only if an ENDO dongle is present.

Start-up ENDO menu

The ENDO menu is started-up through:

▪ taking the ENDO motor out of the dentist element (setting in user menu, option 5).

▪ activating the (M) "Additional motor drives" key afterwards.

S1

S2

S3

S4

Key

Settings

S1

Saves the parameters
Selects the ENDO menu

S2

Changes the value

S3

Changes the value

S4

Moves the flashing cursor
Torque mode, speed, torque, torque mode, .. (ring counter)

▶ Press the "S1" key for an extended time to select the ENDO settings menu. The individual parame‐

ters can be set and changed in this menu.

Setting the handpieces

15

Summary of Contents for ESTETICA E30

Page 1: ...Short instructions for use ESTETICA E30...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...the timer 12 5 1 Select the timer time 12 5 2 Set the timer time 12 6 Setting the handpieces 13 6 1 Adjusting functions 13 6 2 Turbine 14 6 3 INTRA LUX Motor KL 701 KL 703 LED 14 6 4 KL 703 LED in EN...

Page 4: ...very 1 1 Dentist element TM table F E A B C D A Group of keys for the dental chair D Group of keys for the menu B Group of keys for illumination E Group of keys for hygiene C Group of keys for the han...

Page 5: ...dental chair 1 4 Groups of keys Group of keys for the dental chair The keys of the assistant unit each have two functions and show two symbols Assistant unit key Dentist unit key Labelling Chair up ke...

Page 6: ...Operating light Dentist element and assistant el ement Operating light dimming key Dentist element Key X ray viewer Dentist element Direction of motor rotation but ton Dentist element Preselected spr...

Page 7: ...DROclean key Assistant element unassigned key Group of keys for the menu Group of keys for the menu Menu function selection keys Display Group of keys for the timer Key Name Control element Remote Con...

Page 8: ...ives dental chair to rinsing position optional Sets the preset blown air at the dental instru ment does not apply to PiezoLED Chair position direction of motor rotation 4 way switch Changes the positi...

Page 9: ...t Briefly press the desired button Chair automatically moves to the saved position Upon arrival at the saved position the display diode on the button is turned on Saving the automatic position through...

Page 10: ...chair position is saved Recalling an automatic position through the foot control Press the SP foot switch or Press the LP foot switch The chair moves into the selected automatic position Saving an aut...

Page 11: ...display mode for time of day and date Time of day only Time of day without seconds Time of day and date Date only 9 LCD Set contrast of LCD display 10 Language Set menu language Deutsch English Italia...

Page 12: ...ess S4 button 2 level is displayed Users 4 6 can be selected Press button S1 S2 S3 Users 4 5 or 6 will be selected 3 2 Selecting and setting an option S1 S2 S3 S4 Press the Increase value and Decrease...

Page 13: ...o the stand by menu upon closing of the handpiece and endo menus S1 S2 S3 S4 Press the Next key S4 to start up the user menu The user menu displays options and parameters These can be adjusted and cha...

Page 14: ...2 Set the timer time Press the Timer key for an extended period of time to start the timer programming mode The pro gramming mode can be started in the stand by menu only S1 S2 S3 S4 Key Settings S1 S...

Page 15: ...ion Foot control control keys LED control keys Spray status Foot control control keys LED control keys Saving the values Press the Save key to save the values You can save after setting each value or...

Page 16: ...on The direction of motor rotation is displayed on the dentist element Setting the cooling level Remove the handpiece such as turbine motor from the holder The handpiece is active Press the Preselecte...

Page 17: ...s started up through taking the ENDO motor out of the dentist element setting in user menu option 5 activating the M Additional motor drives key afterwards S1 S2 S3 S4 Key Settings S1 Saves the parame...

Page 18: ...ratio 3 1 0 4 to 8 0 Ncm The following parameters can be changed Speed Torque Torque mode Speed The speed can be adjusted as follows in steps of ten in the range from 100 rpm to 500 rpm in steps of f...

Page 19: ...follows in steps of 0 05 Ncm in the range from 0 4 Ncm to 8 Ncm in steps of 2 in the range from 1 to 100 Torque mode Three different torque modes are available Torque only Autoreverse Autoreverse Forw...

Page 20: ...eselect counterclockwise motor rotation Take motor off the holder The handpiece is active Slide the cross switch upward The direction of motor rotation is reversed each time the cross switch is actuat...

Page 21: ...on is available only with the water bottle kit Treatment water that comes into contact with the patient tumbler dental instruments can be disin fected manually or intensively disinfected manually Duri...

Page 22: ...0 8 4 Conditioning the suction system Note Clean the suction hoses after each treatment and disinfect them with DEKASEPTOL Gel daily Application of DEKASEPTOL Gel Note Regarding the air jet suction sy...

Page 23: ...llow it to act With without air jet suction system Venturi Place the suction hoses in the holder DEKASEPTOL Gel Basis Set DEKASEPTOL Gel replacement canister Mat no 1 000 7204 Mat no 1 000 7205 8 5 Re...

Page 24: ...ediately in order to prevent discolouration of the surfaces Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility Guide the tube through the orifice of the coarse sieve of th...

Page 25: ...dosing facility nose is situated right above the opening of the intensive disinfection bottle Then fill up the intensive disinfection bottle with tap water Screw the intensive disinfection bottle tig...

Page 26: ...ssistant element LED on dentist element Bracket on the foot control Assistant element Backrest Kickplate Seat For right left conversion with bench removed To deactivate an activated safety shutoff rem...

Page 27: ...ist Operating device no function Bus hardware error Turn the device off and on Call the service technician to look into the problem if it con tinues to exist Turbine making loud running noises Turbine...

Page 28: ...The suction hoses do not have any suction Slides on the conical sections are closed Open the slide valve Sieves in suction connector are blocked Replace sieves Base plate for vacu stop has been activ...

Page 29: ...isplay shows ID 65 Safety switch of bowl suction has been reached Turn external suction on Check and clean if required the bowl valve Display shows ID 68 Call for service Have a service performed Call...