KaVo ESTETICA E30 Short Instructions For Use Download Page 27

9.2

   

Troubleshooting

Malfunction

Cause

Remedy

Nothing works.

Main switch is off.

▶ Turn on the main switch.

Main service fuse interrupted the

electric circuit.

▶ Unplug the unit from the

mains.

▶ Check and replace, if re‐

quired, the main service fuse.

The main service fuse is situ‐

ated next to the master

switch.

▶ For this purpose, open the

bayonet closure with a screw‐

driver and replace the fine-

wire fuse (T 6.3 H

Mat. no.0.223.2783).

▶ The re-close the bayonet clo‐

sure with the screwdriver.

The patient chair does not

move.

The safety shutoff is activated.
(The LED on the control panel

flashes.)

▶ Check the safety shutoff and

eliminate the reason for the

shutoff.

Display without indicator.

Bus / hardware error.

▶ Turn the device off and on.

▶ Call the service technician to

look into the problem if it con‐

tinues to exist.

Operating device no function.

Bus / hardware error.

▶ Turn the device off and on.

▶ Call the service technician to

look into the problem if it con‐

tinues to exist.

Turbine making loud running

noises.

Turbine wheel faulty.

▶ Replace turbine wheels.

Follow the operating instruc‐

tions for the turbine.

The Satelec Mini LED does not

work.

Also refer to: Instructions for use

for the Satelec Mini LED

See also: Instructions for use for

the Satelec Mini LED

No cold light in the instruments. Cold light not preselected.

▶ Preselect cold light.

The high-pressure lamp or Multi

LED on the instrument is defec‐

tive.

▶ Replace the high-pressure

lamp or Multi LED.

See also: Operating Instructions

of the instrument

No spray in the handpieces.

No spray preselected.

▶ Preselect spray.

Check setting on table-side

butterfly valves

Close the ring for controlling the

spray on the handpieces.

▶ Open the ring for controlling

the spray on the handpieces.

Check setting on table-side

butterfly valves

Troubleshooting / warning signals

25

Summary of Contents for ESTETICA E30

Page 1: ...Short instructions for use ESTETICA E30...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...the timer 12 5 1 Select the timer time 12 5 2 Set the timer time 12 6 Setting the handpieces 13 6 1 Adjusting functions 13 6 2 Turbine 14 6 3 INTRA LUX Motor KL 701 KL 703 LED 14 6 4 KL 703 LED in EN...

Page 4: ...very 1 1 Dentist element TM table F E A B C D A Group of keys for the dental chair D Group of keys for the menu B Group of keys for illumination E Group of keys for hygiene C Group of keys for the han...

Page 5: ...dental chair 1 4 Groups of keys Group of keys for the dental chair The keys of the assistant unit each have two functions and show two symbols Assistant unit key Dentist unit key Labelling Chair up ke...

Page 6: ...Operating light Dentist element and assistant el ement Operating light dimming key Dentist element Key X ray viewer Dentist element Direction of motor rotation but ton Dentist element Preselected spr...

Page 7: ...DROclean key Assistant element unassigned key Group of keys for the menu Group of keys for the menu Menu function selection keys Display Group of keys for the timer Key Name Control element Remote Con...

Page 8: ...ives dental chair to rinsing position optional Sets the preset blown air at the dental instru ment does not apply to PiezoLED Chair position direction of motor rotation 4 way switch Changes the positi...

Page 9: ...t Briefly press the desired button Chair automatically moves to the saved position Upon arrival at the saved position the display diode on the button is turned on Saving the automatic position through...

Page 10: ...chair position is saved Recalling an automatic position through the foot control Press the SP foot switch or Press the LP foot switch The chair moves into the selected automatic position Saving an aut...

Page 11: ...display mode for time of day and date Time of day only Time of day without seconds Time of day and date Date only 9 LCD Set contrast of LCD display 10 Language Set menu language Deutsch English Italia...

Page 12: ...ess S4 button 2 level is displayed Users 4 6 can be selected Press button S1 S2 S3 Users 4 5 or 6 will be selected 3 2 Selecting and setting an option S1 S2 S3 S4 Press the Increase value and Decrease...

Page 13: ...o the stand by menu upon closing of the handpiece and endo menus S1 S2 S3 S4 Press the Next key S4 to start up the user menu The user menu displays options and parameters These can be adjusted and cha...

Page 14: ...2 Set the timer time Press the Timer key for an extended period of time to start the timer programming mode The pro gramming mode can be started in the stand by menu only S1 S2 S3 S4 Key Settings S1 S...

Page 15: ...ion Foot control control keys LED control keys Spray status Foot control control keys LED control keys Saving the values Press the Save key to save the values You can save after setting each value or...

Page 16: ...on The direction of motor rotation is displayed on the dentist element Setting the cooling level Remove the handpiece such as turbine motor from the holder The handpiece is active Press the Preselecte...

Page 17: ...s started up through taking the ENDO motor out of the dentist element setting in user menu option 5 activating the M Additional motor drives key afterwards S1 S2 S3 S4 Key Settings S1 Saves the parame...

Page 18: ...ratio 3 1 0 4 to 8 0 Ncm The following parameters can be changed Speed Torque Torque mode Speed The speed can be adjusted as follows in steps of ten in the range from 100 rpm to 500 rpm in steps of f...

Page 19: ...follows in steps of 0 05 Ncm in the range from 0 4 Ncm to 8 Ncm in steps of 2 in the range from 1 to 100 Torque mode Three different torque modes are available Torque only Autoreverse Autoreverse Forw...

Page 20: ...eselect counterclockwise motor rotation Take motor off the holder The handpiece is active Slide the cross switch upward The direction of motor rotation is reversed each time the cross switch is actuat...

Page 21: ...on is available only with the water bottle kit Treatment water that comes into contact with the patient tumbler dental instruments can be disin fected manually or intensively disinfected manually Duri...

Page 22: ...0 8 4 Conditioning the suction system Note Clean the suction hoses after each treatment and disinfect them with DEKASEPTOL Gel daily Application of DEKASEPTOL Gel Note Regarding the air jet suction sy...

Page 23: ...llow it to act With without air jet suction system Venturi Place the suction hoses in the holder DEKASEPTOL Gel Basis Set DEKASEPTOL Gel replacement canister Mat no 1 000 7204 Mat no 1 000 7205 8 5 Re...

Page 24: ...ediately in order to prevent discolouration of the surfaces Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility Guide the tube through the orifice of the coarse sieve of th...

Page 25: ...dosing facility nose is situated right above the opening of the intensive disinfection bottle Then fill up the intensive disinfection bottle with tap water Screw the intensive disinfection bottle tig...

Page 26: ...ssistant element LED on dentist element Bracket on the foot control Assistant element Backrest Kickplate Seat For right left conversion with bench removed To deactivate an activated safety shutoff rem...

Page 27: ...ist Operating device no function Bus hardware error Turn the device off and on Call the service technician to look into the problem if it con tinues to exist Turbine making loud running noises Turbine...

Page 28: ...The suction hoses do not have any suction Slides on the conical sections are closed Open the slide valve Sieves in suction connector are blocked Replace sieves Base plate for vacu stop has been activ...

Page 29: ...isplay shows ID 65 Safety switch of bowl suction has been reached Turn external suction on Check and clean if required the bowl valve Display shows ID 68 Call for service Have a service performed Call...

Page 30: ......

Page 31: ...1 009 6856 fk 20120703 01 en...

Reviews: