![KaVo ELECTROmatic C Instructions For Use Manual Download Page 73](http://html1.mh-extra.com/html/kavo/electromatic-c/electromatic-c_instructions-for-use-manual_658132073.webp)
Instructions for use ELECTROmatic M/C and PM/PC
12 Information about electromagnetic compatibility | 12.3 Guidelines and manufacturer's statement - Electromag-
netic immunity
Interference im-
munity tests
IEC 60601 test
levels
Compliance level
Electromagnetic environment
- Guidelines
Electrostatic discharge
(ESD) according to
IEC 61000-4-2
± 8 kV contact dis-
charge
± 15 kV atmospheric
discharge
± 8 kV contact dis-
charge
± 15 kV atmospheric
discharge
Floors should be made of wood
or concrete or be fitted with
ceramic tiles. If the floor is fitted
with synthetic material, the relat-
ive humidity must be at least 30
%.
Fast transient elec-
trical interference /
bursts according to
IEC 61000-4-4
± 2 kV for power lines
± 2 kV for input and
output lines
± 2 kV for power lines
and input and output
lines
The quality of the supply voltage
should correspond to that of a
typical business or hospital envir-
onment.
Surges according to
IEC 61000-4-5
± 1 kV push-pull
voltage
± 2 kV common mode
voltage
± 1 kV push-pull
voltage
± 2 kV common mode
voltage
The quality of the supply voltage
should correspond to that of a
typical business or hospital envir-
onment.
Voltage interruptions,
short-term interrup-
tions, and fluctuations
of the supply voltage
according to IEC
61000-4-11
0 % / 0.5 cycles
at 0º to 315º in 45º
increments
0 % / 1 cycle
70 % / 25 cycles
0 % / 250 cycles
0 % / 0.5 cycles
at 0º to 315º in 45º
increments
0 % / 1 cycle
70 % / 25 cycles
0 % / 250 cycles
The quality of the supply voltage
should correspond to that of a
typical business or hospital envir-
onment.
Magnetic field at a
supply frequency
(50/60 Hz) according
to IEC 61000-4-8
30 A/m
30 A/m
Magnetic fields at the mains fre-
quency should correspond to the
typical values in a business and
hospital environment.
12.3 Guidelines and manufacturer's statement -
Electromagnetic immunity
The ELECTROmatic is intended for use in an environment of the kind specified
below. The customer or user of the ELECTROmatic should make sure that the
device is used in an environment of the specified type.
73 / 78
Summary of Contents for ELECTROmatic C
Page 1: ...Instructions for use ELECTROmatic M C and PM PC...
Page 76: ......
Page 77: ......
Page 78: ...1 012 6295 bd 20200110 04 en...