Instructions for use ELECTROmatic M/C and PM/PC
5 Startup | 5.4 Making the device settings
▶
Place the airflow measuring tube
①
(
Mat. no. 0.411.4441
) on the motor.
▶
Press the foot control to start the motor.
▶
Adapt the system pressure of the treatment centre appropriately such that
the cooling airflow is 6 to 9 Nl/min (upper edge of sphere
②
).
▶
Comply with the limits for the system pressure from the technical specifica-
tions for media.
See also:
2
3.6 Technical specifications of the ELECTROmatic, Page 24
5.4 Making the device settings
▶
Press both the plus and minus key for 2 seconds to start the "Settings"
menu.
▶
Press the arrow keys to select the various parameters.
▶
Press the Plus key to increase a value.
or
▶
Press the Minus key to decrease in value.
ð
The settings are always saved instantaneously.
▶
Press both the plus and minus key for 2 seconds to exit from the "Settings"
menu.
The following device settings can be changed according to need and/or dis-
played when the device is being put into service.
Display information
Setting
Displays the software versions.