Instructions for use ELECTROmatic M/C and PM/PC
6 Operation | 6.1 Switching the ELECTROmatic on/off
54 / 78
6 Operation
CAUTION
Germ formation.
Infections.
▶
Before treating a patient, let the spray air and spray water exit for at least
20 seconds.
▶
Before the first start-up and after downtimes (weekends, holidays, vaca-
tions, etc.), the air and water lines must be purged and/or rinsed.
▶
The germ reduction of the ELECTROmatic must be carried out via the
dental unit.
See also:
2
8 Reprocessing steps in accordance with DIN EN ISO 17664, Page 61
NOTICE
Incorrectly set parameters.
Property damage from incorrect input values.
▶
Check all input values on the display before use.
NOTICE
Contaminated or moist compressed air at the compressed air connec-
tion
Premature wear
▶
Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
Note
Please note the transmission/reduction ratio of the attachment handpieces as
these have an impact on the displayed speed of the clamped tool.
6.1 Switching the ELECTROmatic on/off
The unit is ready for operation as soon as it is connected.
▶
To switch the product on, connect the device to the power circuit.
ð
An LED on the rear of the ELECTROmatic M/C flashes.
ð
The treatment parameters are displayed on the display of the
ELECTROmatic PM/PC:
▶
To switch the device off, unplug the unit from the mains.
ð
The LED on the rear of the ELECTROmatic M/C is off.
Summary of Contents for ELECTROmatic C
Page 1: ...Instructions for use ELECTROmatic M C and PM PC...
Page 76: ......
Page 77: ......
Page 78: ...1 012 6295 bd 20200110 04 en...