Instructions for use DEXIS CariVu 3-in-1 by KaVo
7 Processing steps in accordance with ISO 17664 | 7.3 Disassembly
38 / 50
7.3 Disassembly
WARNING
Incomplete processing.
Infection hazard.
▶
To ensure complete processing of all parts, camera, tip and fluorescence
cover need to be removed before processing.
7.4 Manual processing
NOTICE
Never process the device in an ultrasonic cleaner.
Malfunction and material damage.
7.4.1 Manual external cleaning
▶
Clean the DEXIS CariVu 3-in-1 by KaVo, the tip and the fluorescence cover
for at least 1 minute per item under running potable water with an flow rate
2l/min and a temperature of 30 °C ± 5 °C (86 °F ± 9 °F) or until the part is
visually clean.
▶
Clean gently, since excessive force may damage the camera.
▶
Do not bend the silicone legs of the tip apart during manual cleaning.
▶
Visually inspect if the part looks clean and inspect wear and tear. If the part
is not clean, repeat the relevant previous pre-cleaning steps. Worn down
parts should be repaired or replaced.
7.4.2 Manual internal cleaning
This product is not designed for manual internal cleaning.
7.4.3 Manual external disinfection
NOTICE
Never process the device in an ultrasonic cleaner.
Malfunction and material damage.
KaVo recommends the following products based on material compatibility. The
microbiological efficacy must be ensured by the disinfectant manufacturer and
proven by an expert opinion.
Approved disinfectants:
▪ CaviWipes and CaviCide made by Metrex (intermediate disinfection)
Summary of Contents for DEXIS CariVu 3-in-1
Page 1: ...Instructions for use DEXIS CariVu 3 in 1 by KaVo...
Page 49: ......