Instructions for use DEXIS CariVu 3-in-1 by KaVo

6 Operation | 6.4 Recording images

30 / 50

6.4.2 Fluorescence mode

 CAUTION

Possibly hazardous optical radiation emitted from product.

Risk of injury.
Do not stare into the active light source.

 CAUTION

The camera can reach a temperature of up to 48 °C in fluorescence
mode.

Risk of injury.

Avoid too long operating times.

 CAUTION

UV emitted from this product.

In Fluorescence mode, it is recommended to plug the fluorescence cover onto
the camera to be able to record optimally illuminated images.

In this mode, the DEXIS CariVu 3-in-1 by KaVo camera records images in the
fluorescence range for the detection of occlusal caries.
For homogeneous illumination, LEDs are provided at the lens system, which are
blue in this mode.

This light excites the dental hard substance and the metabolic products of cario-
genic bacteria. These substances emit different colours (fluorescence), which al-
lows the user to detect carious activity. This mode can be used to record im-
ages and videos in the fluorescence range for the diagnosis of occlusal caries
and plaque. Like in Intraoral mode, the permanent autofocus assures the cap-
ture of permanently focused images.

Note

It is recommended to switch the light of the treatment unit off in this mode.

Summary of Contents for DEXIS CariVu 3-in-1

Page 1: ...Instructions for use DEXIS CariVu 3 in 1 by KaVo...

Page 2: ...Distributed by DEXIS LLC 1910 North Penn Road Hatfield PA 19440 USA Phone 1 888 883 3947 www dexis com Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...17 3 4 Technical Specifications 17 3 5 Delivery content 19 3 6 System components 20 3 7 Control panel 21 4 Installation of the wall mount 22 5 Startup and shut down 24 5 1 Recommended PC requirements...

Page 4: ...disinfection 40 7 5 3 Automated drying 41 7 6 Care products and systems Servicing 41 7 7 Sterilization 41 7 8 Storage 42 8 Tools and consumables 43 9 Information concerning the electromagnetic compat...

Page 5: ...umber of the product in all inquiries For repairs please contact KaVo Repair Service For scheduling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 F...

Page 6: ...neral marks and symbols See Chapter on User Instructions Hazard Levels Important information for users and service technicians Action request CE mark European Community A product bearing this mark mee...

Page 7: ...safety notes in this document must be observed to prevent personal injury and property damage The warning notes are designated as shown below DANGER In cases which if not prevented directly lead to de...

Page 8: ...have arisen only after the time of delivery in accordance with the General German Freight Forwarders Terms and Conditions Art 28 Outside Germany Note KaVo shall not be held liable for damage arising...

Page 9: ...th respect to compensation are excluded In the event of default and gross negli gence or intent this shall only apply in the absence of mandatory legal regula tions to the contrary KaVo shall not be h...

Page 10: ...g is effective Process the product and accessories appropriately before disposal 2 2 Technical condition A damaged device or components can injure patients users and third parties Visually inspect the...

Page 11: ...ds the national and re gional regulations Only employ the device if the user has the appropriate medical training 2 5 Electrical shock Damage or connection to external devices can lead to electrical s...

Page 12: ...itute a product modification that leads to the loss of CE conformity In the event of damage the responsibility is with the service company or the operator The introduction into the market of a modifie...

Page 13: ...equipped with an autofocus mechanism which will provide a sharp representation of the tooth which is placed in the center of the image The autofocus provides the ability to have sharp images from clo...

Page 14: ...le in two similar models The differ ences between the models are a wall mounted version to store the device with a cable length of 2 5 meters model 1 011 4444 and a table holder version to store the d...

Page 15: ...restricts this device to sale by or on the order of a healthcare professional dentist For dental use only Proper Use According to these regulations this product may only be used for the described appl...

Page 16: ...note the instructions for use Please note the instructions for use HIBC Code Certification CE mark according to Medical Devices Directive EC 93 42 VDE mark MET mark Medical device labeling of medical...

Page 17: ...013 1700 Cable length 1 30 m 3 4 Technical Specifications Complete system Power consumption max 0 5 A Supply voltage 5 V Cable length standard 2 5 m Weight 190 g Length standard 255 mm Diameters appro...

Page 18: ...re 10 to 30 o C Air pressure 800 to 1060 hPa Rel humidity 5 to 95 non condensing Operating altitude max 2000 m Operating environment Permissible installation sites Indoor premises portable Pollution d...

Page 19: ...le holder Camera Wall mount Control panel Sheath 100 pieces The USB cable is used to connect the camera to the computer and to send the captured image video to the imaging software The DEXIS CariVu 3...

Page 20: ...er Fluorescence Cover is used in Fluorescence mode See also 2 6 4 2 Fluorescence mode Page 30 Sheaths The sheaths are used for infection prevention Table holder The table holder is to be placed on the...

Page 21: ...cording Status LED indicating the current status of the camera Blue Standby mode White Camera is active Red Error Buttons for navigation in the imag ing software e g numbered tooth chart if available...

Page 22: ...osition using the wall mount Drill holes with a diameter of 6 mm insert the dowels and screw the wall mount tight with screws Gluing the wall mount to a wall NOTICE No secure hold of camera and wall m...

Page 23: ...Instructions for use DEXIS CariVu 3 in 1 by KaVo 4 Installation of the wall mount Slide the wall mount onto the mounting rail and place the DEXIS CariVu 3 in 1 by KaVo in the wall mount 23 50...

Page 24: ...8 GB memory is recommended Hard Disk 16 GB of free disk space is rec ommended 16 GB of free disk space is rec ommended Display Full HD 1920 1080 or higher Full HD 1920 1080 or higher e g 1920 1200 Gra...

Page 25: ...the control panel Removal of the camera from the table holder or wall mount 6 2 Use a sheath Apply the sheath WARNING Risk of infection due to soiled or contaminated camera Infection Always use the ca...

Page 26: ...d exactly over the ex posure window of the camera This enables sharp imaging Tighten the clear view window in front of the exposure window of the cam era Note In event of bad image quality or milky di...

Page 27: ...p can lead to damage on the camera and the tip Make sure that the contours of the camera align with those of the tip Do not force the tip to snap in Do not slide the tip in axial direction Before the...

Page 28: ...ress off the tip at the predetermined recess and do not touch it by the soft silicone legs 6 4 Recording images 6 4 1 Intraoral mode CAUTION Possibly hazardous optical radiation emitted from product R...

Page 29: ...heath first before use the camera See also 2 6 2 Use a sheath Page 25 The recording window of the imaging software shows a live image Select Intraoral mode still image or video in the imaging software...

Page 30: ...s In this mode the DEXIS CariVu 3 in 1 by KaVo camera records images in the fluorescence range for the detection of occlusal caries For homogeneous illumination LEDs are provided at the lens system wh...

Page 31: ...e a sheath Page 25 To be able to capture an optimum image the fluorescence cover needs to be fully plugged onto the camera This prevents the ingress of extraneous light and is an additional aid for si...

Page 32: ...screen In this mode the DEXIS CariVu 3 in 1 by KaVo camera records images in tran sillumination for the detection of occlusal caries proximal caries secondary caries and cracks Proximal caries Occlusa...

Page 33: ...tip The recording window of the imaging software shows a live image Select an image detail The camera has a fixed focus in this mode Press one of the two trigger buttons The camera switches to Freeze...

Page 34: ...ove the camera from the camera holder Select Fluorescence mode still image or video in the imaging software or with the menu button Select an image detail The camera automatically focuses the selected...

Page 35: ...ws the Intraoral mode exclusively Select an image detail The camera automatically focuses the selected image detail permanent autofocus Press one of the two trigger buttons The camera switches to Free...

Page 36: ...zed by the camera Position the camera to get live video of the selected tooth Look for similar horizontal alignment of the tooth by moving the device hor izontally Adjust the device alternately mesial...

Page 37: ...es No No No Wall Mount Vision Full HD Multi Person use Yes Yes No No No Sheath Vision Full HD Single Use No No No No No 7 2 Preparations at the site of use WARNING Hazard from contaminated products Co...

Page 38: ...F or until the part is visually clean Clean gently since excessive force may damage the camera Do not bend the silicone legs of the tip apart during manual cleaning Visually inspect if the part looks...

Page 39: ...n the medical device and allow the disinfectant to act in accordance with the instructions of the disinfectant manufacturer Comply with the instructions for use of the disinfectant Do not bend the sil...

Page 40: ...which are operated using alkaline cleaning agents The validations were conducted with the VARIO TD program the cleaning agent neodisher MediClean forte and the neutralisation agent neodisher Z For pro...

Page 41: ...age must be large enough for the product so that the packaging is not stretched The quality and use of the sterilization packaging must comply with applica ble standards and be suitable for the steril...

Page 42: ...er using the gravity method 10 minutes at 135 C 275 F Drying time 30 min Use in accordance with the manufacturer s Instructions for Use 7 8 Storage Processed products must be stored protected from bac...

Page 43: ...hich illuminates the tooth from the sides The Tip Vision Full HD is fully repro cessable The Tip Vision Full HD is provided in a 1 pc package as well as in a 4 pcs package 1 013 5709 Fluorescence Cove...

Page 44: ...Vision Full HD has a back plate which can be fixed to a surface by using the at tached tape or by using screws The main body is also equipped with a magnet inside This magnet can be detected by the m...

Page 45: ...mission of the device is very low and interference with adja cent electronic devices is unlikely HF emissions according to CISPR 11 Class B The DEXIS CariVu 3 in 1 by KaVo is suitable for use in all e...

Page 46: ...ccording to EN 61000 4 2 8 kV contact discharge 15 kV atmospheric discharge 8 kV contact discharge 2 4 8 15 kV atmo spheric discharge Floors should be made of wood or concrete or be fitted with cerami...

Page 47: ...stationary wireless radio transmitters as measured locallyc should be lower than the conformance level at all frequencies d Interference is possible in the vicinity of devices bearing the following ic...

Page 48: ...XIS CariVu 3 in 1 by KaVo is intended for use in an electromagnetic en vironment in which the HF interference parameters are controlled The cus tomer or the user of the DEXIS CariVu 3 in 1 by KaVo can...

Page 49: ......

Page 50: ...1 013 7129 bd 20220323 06 en US...

Reviews: