CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisa‐
tion cycle!
The KaVo medical device has a maximum temperature resistance up to 138 ℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the fol‐
lowing sterilisation processes:
▪ Autoclave with three times pre-vacuum:
– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Autoclave using the gravity method:
– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F) or
alternatively
– at least 90 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4 °F)
▶ Use according to the manufacturer's Instructions for Use.
7.8
Storage
▶ Reprocessed products should be stored protected from dust with minimum ex‐
posure to germs in a dry, dark and cool space.
▶ Comply with the expiry date of the sterilised items.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830
7 Reprocessing methods according to ISO 17664 | 7.8 Storage
20/22