3.2
Technical data
Minimum pressure
2.1 bar (30 psi)
Drive pressure
2.1 to 2.3 bar (30 to 33 psi)
Recommendation
2.1 bar (30 psi)
Air consumption
35 to 40 Nl/min (1.24 to 1.41 cfm)
Idle speed
350,000 to 400,000 rpm
Recommended operating pressure
2 to 3 N (0.45 to 0.68 lbs)
COMPACTtorque 636 turbines can be attached to all hoses with Borden 2-hole
connections.
COMPACTtorque 636 turbines can be attached to all hoses with 4-hole connec‐
tions.
3.3
Transportation and storage conditions
CAUTION
It is hazardous to start up the medical device after it has been stored strongly
refrigerated.
This can cause the medical device to malfunction.
▶ Prior to start-up, very cold products must be heated to a temperature of 20°C
to 25°C (68°F to 77°F).
Temperature: -20°C to +70°C (-4°F to +158°F)
Relative humidity: 5% RH to 95% RH absence of condensation
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Protect from moisture
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830
3 Product description | 3.2 Technical data
6/22