1.010.0216 · sö · 20170120 - 2 · en

Summary of Contents for 1.011.6740

Page 1: ...Instructions for use SMARTmatic PROPHY S19 1 011 6740 SMARTmatic PROPHY S19 K 1 011 6741...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...ngs 16 7 Reprocessing steps in accordance with ISO 17664 17 7 1 Preparations at the site of use 17 7 2 Cleaning 17 7 2 1 Manual external cleaning 17 7 2 2 Automated external cleaning 17 7 2 3 Manual i...

Page 4: ...avo com Target group This document is intended for dentists and their assistants The startup section is also intended for service technicians General marks and symbols Refer to the chapter on Safety W...

Page 5: ...PROPHY S19 1 011 6740 SMARTmatic PROPHY S19 K 1 011 6741 1 User instructions CAUTION In cases which if not prevented could lead to minor or moderate injury NOTICE In cases which if not prevented coul...

Page 6: ...In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air Use gloves or finger guard whenever you test insert and remove the tool 2 2...

Page 7: ...acility for changing the speed and the direction of rotation must be present The medical device may only be combined with a treatment centre released by KaVo Comply with the Instructions for Use of th...

Page 8: ...19 K 1 011 6741 2 Safety 2 5 Service and repair 8 28 KaVo recommends specifying in house service intervals where the medical device is brought to a professional shop for cleaning servicing and a funct...

Page 9: ...ool to support the periodontal treatment by polish ing with a rotating tool A medical device according to relevant national statutory regulations Proper use According to these provisions this medical...

Page 10: ...mission 8 1 Cooling air flow 5 5 to 9 5 Nl min Prophylaxis heads with a Doriot connector can be inserted Shaft and or drive shaft in accordance with DIN EN ISO 14457 for metallic and ceramic materials...

Page 11: ...Transportation and storage conditions NOTICE Startup after refrigerated storage Malfunction Prior to startup very cold products must be heated up to a temperature of 20 C to 25 C 68 F to 77 F Temperat...

Page 12: ...Dispose of the product in appropriate manner Infection hazard Before disposal reprocess and sterilise the product and accessories appropri ately NOTICE Damage from soiled and moist cooling air Contami...

Page 13: ...to ensure that it is se curely locked onto the motor coupling NOTICE Removing and attaching the handpiece while the drive motor is rotating Damage to the driver Never attach or remove the handpiece w...

Page 14: ...is heads Injury to the patient or damage to the medical device Comply with the instructions for use and the intended use of the tools Only use prophylaxis heads that do not deviate from the specified...

Page 15: ...the prophylaxis head 5 4 Removing the prophylaxis head Remove the prophylaxis head axially from the handpiece CAUTION Hazard due to incorrectly stored instrument Injury and infection from clamped prop...

Page 16: ...is irregularly heated The medical device is too hot while working Service the medical device When the speed drops or is uneven Service the medical device An O ring is missing on the motor coupling Re...

Page 17: ...lly or in a washer disinfector only 7 2 1 Manual external cleaning Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Brush e g medium hard toothbrush Brush off under flowing tap water 7 2 2 Au...

Page 18: ...line cleaning agents The validations were conducted with the VARIO TD program the cleaning agent neodisher MediClean and the neutralisation agent neodisher Z For program settings as well as cleansers...

Page 19: ...external and internal disinfection KaVo recommends washer disinfectors in compliance with EN ISO 15883 1 that are operated with alkaline cleaning agents The validations were conducted with the VARIO T...

Page 20: ...rvicing with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool Cover the pr...

Page 21: ...xpansion pressure for the interior cleaning of inor ganic residues and optimum care no validated cleaning of the interior in accordance with German RKI requirements KaVo recommends servicing the produ...

Page 22: ...rvice button for at least three seconds until the spray canister control LED flashes three times consecutively The device is in chuck service mode Remove the service coupling of the chuck from the sid...

Page 23: ...to moisture Damage to product Immediately remove the product from the steam steriliser after the sterilisation cycle The KaVo medical device has a maximum temperature resistance up to 138 280 4 F Sele...

Page 24: ...ent stand 2151 0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Spray head INTRA KaVo Spray 0 411 9911 INTRA service coupling 1 009 6143 Material summary Mat No KaVo Spray 21...

Page 25: ...for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcin...

Page 26: ......

Page 27: ......

Page 28: ...1 010 0216 s 20170120 2 en...