Instructions for use SMARTmatic PROPHY S19 - 1.011.6740 SMARTmatic PROPHY S19 K - 1.011.6741
2 Safety | 2.1 Infection hazard
6 / 28
2 Safety
The instructions for use are a component of the product and must be read carefully
prior to use and be accessible at all times.
The device may only be used in accordance with the instructions for use, any other
type of use is not permitted.
2.1 Infection hazard
Patients, users or third parties could be infected by contaminated medical devices.
▶ Take suitable personal protective measures.
▶ Follow the instructions for using the components.
▶ Before the initial startup and after each use, prepare and sterilise the medical
device and accessories accordingly.
▶ Carry out the cleaning and sterilisation as described in the instructions for use.
The procedure has been validated by the manufacturer.
▶ It is essential to ensure the effectiveness of the cleaning and sterilisation in the
case of deviation in procedure.
▶ Prior to disposal, the product and accessories must be appropriately prepared or
sterilised.
▶ In the case of injury to soft tissue, do not continue treatment in the oral cavity with
instruments driven by compressed air.
▶ Use gloves or finger guard whenever you test, insert, and remove the tool.
2.2 Technical condition
A damaged device or components could injure patients, users and third parties.
▶ Only operate devices or components if they are undamaged on the outside.
▶ Check that the device is working properly and is in satisfactory condition before
each use.
▶ Have parts with sites of breakage or surface changes checked by the Service.
▶ If the following defects occur, stop working and have the service personnel carry
out repair work:
▪ Malfunctions
▪ Damage
▪ Irregular running noise
▪ Excessive vibration
▪ Overheating
▪ Tool is not seated firmly in the handpiece
Observe the following instructions in order to guarantee optimum functioning and pre-
vent material damage:
▶ Service the medical device with care products and systems regularly as described
in the instructions for use.
▶ The device should be cleaned, serviced and stored in a dry location, according to
instructions, if it will not be used for a longer period.
High torque of micromotors can lead to severe burn injuries.
▶ Service micromotors regularly.
▶ Do no use any damaged motors.
Summary of Contents for 1.011.6740
Page 1: ...Instructions for use SMARTmatic PROPHY S19 1 011 6740 SMARTmatic PROPHY S19 K 1 011 6741...
Page 26: ......
Page 27: ......
Page 28: ...1 010 0216 s 20170120 2 en...