background image

Instructions for use

SURGmatic S201 XL - 1.010.2332 
SURGmatic S201 XC - 1.010.2333

Summary of Contents for 1.010.2332

Page 1: ...Instructions for use SURGmatic S201 XL 1 010 2332 SURGmatic S201 XC 1 010 2333 ...

Page 2: ...istributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Page 3: ...ubleshooting 15 6 1 Check for malfunctions 15 6 2 Troubleshooting 15 6 2 1 Cleaning the spray clip and the spray tube 15 7 Reprocessing steps in accordance with EN ISO 17664 17 7 1 Preparation at the site of use 17 7 2 Non fixing preliminary cleaning of the spray clip and spray tube 17 7 3 Manual reprocessing 18 7 3 1 Manual internal and external cleaning and internal and external disinfection 18 ...

Page 4: ...scheduling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll free Direct Customer Service 1 888 ASK KAVO 888 275 5286 Email techservice kavokerr com www kavo com Target group The instructions for use are intended for medical professionals in particular dentists and office personnel The section on Commission...

Page 5: ...torage conditions Temperature range Transportation and storage conditions Air pressure Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and material damage The warning notes are designated as shown below DANGER In cases which if not prevented dir...

Page 6: ...Instructions for use SURGmatic S201 XL 1 010 2332 SURGmatic S201 XC 1 010 2333 1 User instructions 6 26 NOTICE In cases which if not prevented could lead to material damage ...

Page 7: ...riately before disposal 2 2 Technical condition A damaged device or components could injure patients users and third parties Only operate devices or components if they are undamaged on the outside Check that the device is working properly and is in satisfactory condition before each use Have parts with sites of breakage or surface changes checked by the Ser vice If the following defects occur stop...

Page 8: ...ed range and the direction of rotation can lead to injury Control facility for changing the speed and the direction of rotation must be present The medical device may only be combined with a treatment centre released by KaVo Comply with the Instructions for Use of the treatment centre 2 4 Qualification of personnel Application of the product by users without the appropriate medical training could ...

Page 9: ...extraction implantology and maxillo facial surgery A medical device according to relevant national statutory regulations Proper use According to these regulations this product may only be used for the described application by a properly trained user You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructions f...

Page 10: ...dium connection The contra angle handpiece can be mounted on all INTRAmatic LUX motors and motors with a connector in accordance with ISO 3964 CAUTION The SURGmatic S201 XL XC contra angle handpiece consists of a shank and a head It is not permissible to combine the unit with other heads shanks 3 3 Transportation and storage conditions Do not store in a refrigerated environment Temperature 20 C to...

Page 11: ...unt of cooling water Thermal damage to the dental pulp Adjust the water amount for the spray cooling to a minimum of 50 ml min CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check the spray water channels and clean the spray tube with the nozzle needle Mat no 0 410 0931 according to need Switch off spray...

Page 12: ...n place it can become detached during treatment Do not mount or remove the head while it is rotating Before each treat ment check that the head is firmly seated and that the clamping ring is tight Rotate the clamping ring in the direction of the arrow until it hits the stop and hold it there Insert the head to the stop Make sure that the catches engage properly Rotate the clamping ring in the dire...

Page 13: ...ly snaps into place Pull on the medical device to make sure that it is securely affixed to the coupling 5 4 Pulling the contra angle handpiece off the motor coupling Unlock the medical device from the motor coupling by twisting it slightly and then pulling it along its axis 5 5 Inserting the milling tool or diamond grinder Note Only use carbide cutters or diamond drill bits that conform to EN ISO ...

Page 14: ...rt or remove the insert to prevent injury and infection Insert the bur or diamond grinder into the segment of the head drive by twisting the tool slightly and push it to the end stop Activate the pushbut ton if necessary Check if the tool is seated securely by pulling on it 5 6 Removing the milling tool or diamond grinder WARNING Hazard from rotating dental bur or diamond grinder Lacerations and d...

Page 15: ...t while working Service the medical device When the speed drops or is uneven Service the medical device An O ring is missing on the motor coupling Replace O ring See also 2 Instructions for use of motor 6 2 Troubleshooting 6 2 1 Cleaning the spray clip and the spray tube CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and dam...

Page 16: ...ions for use SURGmatic S201 XL 1 010 2332 SURGmatic S201 XC 1 010 2333 6 Troubleshooting 6 2 Troubleshooting 16 26 Use the nozzle pin Mat no 0 410 0931 to free the water passage on the spray clip on both sides ...

Page 17: ...od immediately Do not place in solutions or similar substances 7 2 Non fixing preliminary cleaning of the spray clip and spray tube The non fixing preliminary cleaning is a central constituent and must be carried out prior to the automatic reprocessing Accessories required Demineralised water 30 C 2 C 86 F 3 6 F Nozzle pin Brush e g medium hard toothbrush Disposable syringe Note Before cleaning ta...

Page 18: ...leaned in the ultra sonic cleaner 7 3 1 Manual internal and external cleaning and internal and external disinfection Not applicable 7 3 2 Manual drying Blow off the outside and inside with compressed air until water drops are no longer visible Immediately after drying lubricate the KaVo medical device with care agents from the KaVo care system See also 2 7 5 Care products and systems Servicing Pag...

Page 19: ...eaning or disinfection in a thermodisinfector attach the head to an appropriate shank KaVo recommends washer disinfectors according to ISO 15883 1 which are operated using alkaline cleaning agents For programme settings as well as cleansers and disinfectants to be used please refer to the Instructions for Use of the washer disinfector For the spray clip and spray tube also use the adapter for exte...

Page 20: ...over the medical device with the KaVo CLEANpac bag and place it on the corresponding care adapter Press the spray key for 1 to 2 seconds Care of chucking system KaVo recommends cleaning and servicing the chuck system once a week Remove the tool from the medical device Position the tip of the spray nipple in the opening and apply the spray Press the spray key for 1 to 2 seconds 7 5 2 Servicing with...

Page 21: ...e together with the guide bush of the chuck to be ser viced against the tip of the service coupling Press the button marked with the chuck service symbol Note Close the chuck service mode Option 1 Place the dental handpieces in the QUATTROcare PLUS 2124 A close the front door and start theservice procedure Option 2 After three minutes with no service procedure running the device automatically swit...

Page 22: ...uitable process from the following sterilisa tion processes Steriliser with triple pre vacuum at least 3 minutes at 135 C 275 F Drying time 16 min Steriliser using the gravity method at least 10 minutes at 135 C 275 F Drying time 30 min Remove contra angle handpieces and turbine handpieces from the steam steriliser immediately after completion of the sterilisation cycle Use according to the manufa...

Page 23: ...1 7941 Spray clip 1 002 3377 Coupling piece 0 593 0361 Nozzle pin 0 410 0931 Instrument stand 3 005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Only for the USA Material summary Mat no KaVo Spray USA and Canada 2113 A 0 411 9660 QUATTROcare plus Spray USA and Canada 2141 P 1 005 4524 Chuck servicing set 1 003 1253 Canada only Material summary Mat No KaVo Spray Canada 2114 ...

Page 24: ...e for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo s instructions for use or other manufacturer s instru...

Page 25: ......

Page 26: ...1 011 1912 Fk 20190311 02 en ...