Instructions for use SURGmatic S201 XL - 1.010.2332 SURGmatic S201 XC - 1.010.2333
5 Operation | 5.3 Attaching the contra-angle handpiece to motor coupling
5.3 Attaching the contra-angle handpiece to motor
coupling
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked in place can become disconnected
from the motor coupling and fall off.
▶
Carefully pull on the medical device before each treatment to ensure that it
is securely locked onto the motor coupling.
NOTICE
Removing and attaching the straight or contra-angle handpiece while
the drive motor is rotating.
Damage to the driver.
▶
Never attach or remove the straight or contra-angle handpiece while the
drive motor is rotating.
▶
Lightly spray O-rings on motor coupling with KaVo Spray.
▶
Attach the medical device to the motor coupling and turn it until the latch
audibly snaps into place.
▶
Pull on the medical device to make sure that it is securely affixed to the
coupling.
5.4 Pulling the contra-angle handpiece off the motor
coupling
▶
Unlock the medical device from the motor coupling by twisting it slightly
and then pulling it along its axis.
5.5 Inserting the milling tool or diamond grinder
Note
Only use carbide cutters or diamond drill bits that conform to EN ISO 1797
type 1 or EN ISO 17509, are made of steel or hard metal, and meet the fol-
lowing criteria:
- Shaft diameter: 2.334 to 2.350 mm
- Overall length: max. 45 mm
- Blade diameter: max. 10 mm
Using surgical handpieces with hexagon in combination with tools having a
hexagon head socket at the shaft allows deformations at the tool shaft to be
prevented.
It prevents the tool from getting lodged or being difficult to take off.
WARNING
Use of unauthorised cutters or grinders.
Injury to the patient or damage to the medical device.
▶
Observe the instructions for use and use the cutter or grinder properly.
▶
Only use cutters or grinders that do not deviate from the specified data.
13 / 26