Instructions for use SURGmatic S201 XL - 1.010.2332 SURGmatic S201 XC - 1.010.2333
7 Reprocessing steps in accordance with EN ISO 17664 | 7.8 Storage
22 / 26
NOTICE
Contact corrosion due to moisture.
Damage to product.
▶
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
The KaVo medical device has a maximum temperature resistance up to 138
℃
(280.4 °F).
Sterilisation parameters:
Depending on the device, select a suitable process from the following sterilisa-
tion processes.
Steriliser with triple pre-vacuum:
▪ at least 3 minutes at 135 °C (275 °F)
Drying time: 16 min.
Steriliser using the gravity method:
▪ at least 10 minutes at 135°C (275 °F)
Drying time: 30 min.
▶
Remove contra-angle handpieces and turbine handpieces from the steam
steriliser immediately after completion of the sterilisation cycle.
▶
Use according to the manufacturer's Instructions for Use.
7.8 Storage
Prepared products must be stored appropriately in a dry, dark, cool room such
that they are protected from germs (as far as possible) and dust.
Note
Comply with the expiry date of the sterilised items.