KaVo 1.007.0150 Instructions For Use Manual Download Page 7

Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150

2 Safety | 2.1 Infection hazard

7 / 28

2 Safety

NOTE

All serious events occurring in relation to the product must be reported to the
manufacturer and the competent authority of the member state, in which the
user and/or patient resides.

The instructions for use are an integral part of the product and must be read
carefully prior to use and must be accessible at all times. 
The device may only be used in accordance with the intended use, any other
type of use is not permitted.

2.1 Infection hazard

Patients, users or third parties could be infected by contaminated medical
devices.

4

Take suitable personal protective measures.

4

Follow the instructions for use of the components.

4

Before initial startup and after each use, reprocess the product and ac-
cessories appropriately.

4

Carry out the reprocessing as described in the instructions for use. The pro-
cedure has been validated by the manufacturer.

4

If you deviate from this validated procedure, make sure that the repro-
cessing procedure is effective.

4

Reprocess the product and accessories appropriately before disposal.

2.2 Explosion hazard

Electrical sparks in the product can lead to explosion or fire.

4

Do not use the product in explosion hazard areas.

4

Do not operate the product in an oxygen-enriched atmosphere.

4

Do not use the product in the vicinity of flammable gases.

2.3 Electrical shock

Connecting a non-KaVo system to the medical device can lead to electrical
shock and injury to patient, user and third parties.

4

During installation and operation of the product on treatment devices and
equipment from other manufacturers, comply with the provisions of "Protec-
tion from electrical shock," "Leakage current," and "Not grounding the ap-
plied part" in accordance with DIN EN IEC 60601-1.

4

Make sure to combine the product only with a treatment centre / control
unit that has been released by KaVo.

4

Make sure that the product combined with the dental treatment unit / con-
trol unit meets the requirements in accordance with DIN EN IEC 60601-1.

2.4 Improper use

The bluish LED light of the motors can damage the cornea or lens of the eye.

4

Do not gaze into the lamp when it is in operation.

4

Use adequate shielding for eye protection.

2.5 Technical condition

If damaged, the device or components could injure patients, users and third
parties.

4

Only operate devices or components if they show no signs of damage on the
outside.

Summary of Contents for 1.007.0150

Page 1: ...Instructions for use INTRA LUX Motor KL 703 LED 1 007 0150...

Page 2: ...Distribution KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...motor to the supply hose 16 5 2 Regulating the spray water 16 5 3 Attaching an instrument to the motor 16 5 4 Removing an instrument from the motor 17 6 Checking for malfunctions and troubleshooting 1...

Page 4: ...ice If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service einrichtungen kavo com or service treatmentunits kavo com Please refer to the serial num...

Page 5: ...es Transportation and storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from...

Page 6: ...or use INTRA LUX Motor KL 703 LED 1 007 0150 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to m...

Page 7: ...ore disposal 2 2 Explosion hazard Electrical sparks in the product can lead to explosion or fire 4 Do not use the product in explosion hazard areas 4 Do not operate the product in an oxygen enriched a...

Page 8: ...of time 2 6 Electromagnetic fields Electromagnetic fields might interfere with the functions of implanted systems such as pacemakers 4 Ask patients if they have a cardiac pacemaker or other system im...

Page 9: ...als where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As a result of the use of NON...

Page 10: ...fety 2 11 Disposal 10 28 2 11 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further...

Page 11: ...n areas at risk of explosion Intended only for dental treatment by a dental professional the product must not be modified or used for any other purpose since this may be haz ardous A medical device ac...

Page 12: ...011 2600 ELECTROmatic TMM TMC Mat no 0 696 0161 KL 2 Mat no 1 004 4317 KL 3 Mat no 1 006 1037 KL 4 Mat no 1 003 3490 ELECTROtorque TLC Mat no 1 001 8748 ELECTROtorque PLUS Motor voltage rating 22 V AC...

Page 13: ...duct to which the motor is to be attached Check the suitability of the product for operation with the motor NOTE Designed for operation on KaVo units 4 ESTETICA E80 ESTETICA E70 ESTETICA E50 ESTETICA...

Page 14: ...pressure 700 hPa to 1060 hPa 10 psi to 15 psi 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be...

Page 15: ...to 7 Reprocessing steps in accordance with ISO 17664 Page 20 Currently applicable packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employing wa...

Page 16: ...speed is appropriate for the preparation on hand 4 Use the requisite minimal amount of spray water 4 Rotate the regulating sleeve of the motor hose to adjust the water portion of the spray in a contin...

Page 17: ...ions for use INTRA LUX Motor KL 703 LED 1 007 0150 5 Operation 5 4 Removing an instrument from the motor 17 28 5 4 Removing an instrument from the motor 4 Pull the instrument off the motor in axial di...

Page 18: ...n 4 Missing O ring of the motor coupling needs to be replaced 4 If the intensity of the KaVo Mini LED lamp is low increase the cold light in tensity on the unit 4 If the KaVo Mini LED lamp glows red o...

Page 19: ...k properly Case 1 KaVo Mini LED lamp is on Case 2 The KaVo Mini LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 The KaVo Mini LED lamp...

Page 20: ...films may be formed in the lines This is associated with a risk of infection This results in the following requirements for the workflow at the practice 4 Rinse all withdrawal sites of the water condu...

Page 21: ...cal device with a second CaviWipes cleaning and disinfection wipe to remove any remaining invisible contaminations Manual external disinfection 4 Finally thoroughly wipe down the outer surfaces of the...

Page 22: ...rvice regularly with suitable agents CAUTION Excess care products leaking from the contra angle handpiece oil cleanser and disinfectant can penetrate into the motor and cause it to malfunction 4 Pleas...

Page 23: ...the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Autoclave using the gra...

Page 24: ...or KL 703 LED 1 007 0150 8 Consumables 24 28 8 Consumables Available from specialised dental dealers Material summary Mat No KaVo Mini LED lamp 1 007 8474 O ring 8 3x0 68 0 200 6120 Airflow measuring...

Page 25: ...or defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connection instruc...

Page 26: ......

Page 27: ......

Page 28: ...1 008 1048 bd 20220914 06 en...

Reviews: