Instructions for use INTRA LUX Motor KL 703 LED – 1.007.0150

7 Reprocessing steps in accordance with ISO 17664-2 | 7.3 Automated reprocessing

22 / 28

4

If there is any residual moisture after cleaning and disinfecting, soak up the
residual moisture with a soft cotton cloth and then wipe dry.

4

Carefully apply low doses of compressed air to the inside and outside of the
disassembled sleeve until no droplets of water are visible any longer.

7.3 Automated reprocessing

Not applicable.

7.4 Care products and systems - Servicing

The bearing sites of the motors are provided with a high-quality, lifelong lubric-
ation.

 CAUTION

Improper service and care.

Risk of injury.

4

Service regularly with suitable agents.

CAUTION

Excess care products leaking from the contra-angle handpiece (oil, cleanser and
disinfectant) can penetrate into the motor and cause it to malfunction.

4

Please make sure to always follow the care instructions supplied with the in-
struments.

4

Unscrew the sleeve turning it counterclockwise and pull it off from the front
of the motor.

4

Wipe off the inside and outside of the sleeve with a wet cloth once a week.

4

Wipe off the outside of the motor without the sleeve using a dry cloth.

7.5 Packaging

NOTE

The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!

4

Seal the medical device separately in a sterile pack.

Summary of Contents for 1.007.0150

Page 1: ...Instructions for use INTRA LUX Motor KL 703 LED 1 007 0150...

Page 2: ...Distribution KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...motor to the supply hose 16 5 2 Regulating the spray water 16 5 3 Attaching an instrument to the motor 16 5 4 Removing an instrument from the motor 17 6 Checking for malfunctions and troubleshooting 1...

Page 4: ...ice If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service einrichtungen kavo com or service treatmentunits kavo com Please refer to the serial num...

Page 5: ...es Transportation and storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from...

Page 6: ...or use INTRA LUX Motor KL 703 LED 1 007 0150 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to m...

Page 7: ...ore disposal 2 2 Explosion hazard Electrical sparks in the product can lead to explosion or fire 4 Do not use the product in explosion hazard areas 4 Do not operate the product in an oxygen enriched a...

Page 8: ...of time 2 6 Electromagnetic fields Electromagnetic fields might interfere with the functions of implanted systems such as pacemakers 4 Ask patients if they have a cardiac pacemaker or other system im...

Page 9: ...als where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As a result of the use of NON...

Page 10: ...fety 2 11 Disposal 10 28 2 11 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further...

Page 11: ...n areas at risk of explosion Intended only for dental treatment by a dental professional the product must not be modified or used for any other purpose since this may be haz ardous A medical device ac...

Page 12: ...011 2600 ELECTROmatic TMM TMC Mat no 0 696 0161 KL 2 Mat no 1 004 4317 KL 3 Mat no 1 006 1037 KL 4 Mat no 1 003 3490 ELECTROtorque TLC Mat no 1 001 8748 ELECTROtorque PLUS Motor voltage rating 22 V AC...

Page 13: ...duct to which the motor is to be attached Check the suitability of the product for operation with the motor NOTE Designed for operation on KaVo units 4 ESTETICA E80 ESTETICA E70 ESTETICA E50 ESTETICA...

Page 14: ...pressure 700 hPa to 1060 hPa 10 psi to 15 psi 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be...

Page 15: ...to 7 Reprocessing steps in accordance with ISO 17664 Page 20 Currently applicable packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employing wa...

Page 16: ...speed is appropriate for the preparation on hand 4 Use the requisite minimal amount of spray water 4 Rotate the regulating sleeve of the motor hose to adjust the water portion of the spray in a contin...

Page 17: ...ions for use INTRA LUX Motor KL 703 LED 1 007 0150 5 Operation 5 4 Removing an instrument from the motor 17 28 5 4 Removing an instrument from the motor 4 Pull the instrument off the motor in axial di...

Page 18: ...n 4 Missing O ring of the motor coupling needs to be replaced 4 If the intensity of the KaVo Mini LED lamp is low increase the cold light in tensity on the unit 4 If the KaVo Mini LED lamp glows red o...

Page 19: ...k properly Case 1 KaVo Mini LED lamp is on Case 2 The KaVo Mini LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 The KaVo Mini LED lamp...

Page 20: ...films may be formed in the lines This is associated with a risk of infection This results in the following requirements for the workflow at the practice 4 Rinse all withdrawal sites of the water condu...

Page 21: ...cal device with a second CaviWipes cleaning and disinfection wipe to remove any remaining invisible contaminations Manual external disinfection 4 Finally thoroughly wipe down the outer surfaces of the...

Page 22: ...rvice regularly with suitable agents CAUTION Excess care products leaking from the contra angle handpiece oil cleanser and disinfectant can penetrate into the motor and cause it to malfunction 4 Pleas...

Page 23: ...the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Autoclave using the gra...

Page 24: ...or KL 703 LED 1 007 0150 8 Consumables 24 28 8 Consumables Available from specialised dental dealers Material summary Mat No KaVo Mini LED lamp 1 007 8474 O ring 8 3x0 68 0 200 6120 Airflow measuring...

Page 25: ...or defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connection instruc...

Page 26: ......

Page 27: ......

Page 28: ...1 008 1048 bd 20220914 06 en...

Reviews: