Instructions for use

INTRA LUX Motor KL 703 LED – 1.007.0150

Summary of Contents for 1.007.0150

Page 1: ...Instructions for use INTRA LUX Motor KL 703 LED 1 007 0150...

Page 2: ...Distribution KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...motor to the supply hose 16 5 2 Regulating the spray water 16 5 3 Attaching an instrument to the motor 16 5 4 Removing an instrument from the motor 17 6 Checking for malfunctions and troubleshooting 1...

Page 4: ...ice If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service einrichtungen kavo com or service treatmentunits kavo com Please refer to the serial num...

Page 5: ...es Transportation and storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from...

Page 6: ...or use INTRA LUX Motor KL 703 LED 1 007 0150 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to m...

Page 7: ...ore disposal 2 2 Explosion hazard Electrical sparks in the product can lead to explosion or fire 4 Do not use the product in explosion hazard areas 4 Do not operate the product in an oxygen enriched a...

Page 8: ...of time 2 6 Electromagnetic fields Electromagnetic fields might interfere with the functions of implanted systems such as pacemakers 4 Ask patients if they have a cardiac pacemaker or other system im...

Page 9: ...als where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depending on the frequency of use As a result of the use of NON...

Page 10: ...fety 2 11 Disposal 10 28 2 11 Disposal This product is subject to the EC directive governing waste electrical and elec tronic devices and must be submitted to special disposal in Europe 4 For further...

Page 11: ...n areas at risk of explosion Intended only for dental treatment by a dental professional the product must not be modified or used for any other purpose since this may be haz ardous A medical device ac...

Page 12: ...011 2600 ELECTROmatic TMM TMC Mat no 0 696 0161 KL 2 Mat no 1 004 4317 KL 3 Mat no 1 006 1037 KL 4 Mat no 1 003 3490 ELECTROtorque TLC Mat no 1 001 8748 ELECTROtorque PLUS Motor voltage rating 22 V AC...

Page 13: ...duct to which the motor is to be attached Check the suitability of the product for operation with the motor NOTE Designed for operation on KaVo units 4 ESTETICA E80 ESTETICA E70 ESTETICA E50 ESTETICA...

Page 14: ...pressure 700 hPa to 1060 hPa 10 psi to 15 psi 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be...

Page 15: ...to 7 Reprocessing steps in accordance with ISO 17664 Page 20 Currently applicable packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employing wa...

Page 16: ...speed is appropriate for the preparation on hand 4 Use the requisite minimal amount of spray water 4 Rotate the regulating sleeve of the motor hose to adjust the water portion of the spray in a contin...

Page 17: ...ions for use INTRA LUX Motor KL 703 LED 1 007 0150 5 Operation 5 4 Removing an instrument from the motor 17 28 5 4 Removing an instrument from the motor 4 Pull the instrument off the motor in axial di...

Page 18: ...n 4 Missing O ring of the motor coupling needs to be replaced 4 If the intensity of the KaVo Mini LED lamp is low increase the cold light in tensity on the unit 4 If the KaVo Mini LED lamp glows red o...

Page 19: ...k properly Case 1 KaVo Mini LED lamp is on Case 2 The KaVo Mini LED lamp is faint 4 Increase the cold light intensity on the unit until the desired illuminance is reached Case 3 The KaVo Mini LED lamp...

Page 20: ...films may be formed in the lines This is associated with a risk of infection This results in the following requirements for the workflow at the practice 4 Rinse all withdrawal sites of the water condu...

Page 21: ...cal device with a second CaviWipes cleaning and disinfection wipe to remove any remaining invisible contaminations Manual external disinfection 4 Finally thoroughly wipe down the outer surfaces of the...

Page 22: ...rvice regularly with suitable agents CAUTION Excess care products leaking from the contra angle handpiece oil cleanser and disinfectant can penetrate into the motor and cause it to malfunction 4 Pleas...

Page 23: ...the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Autoclave using the gra...

Page 24: ...or KL 703 LED 1 007 0150 8 Consumables 24 28 8 Consumables Available from specialised dental dealers Material summary Mat No KaVo Mini LED lamp 1 007 8474 O ring 8 3x0 68 0 200 6120 Airflow measuring...

Page 25: ...or defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connection instruc...

Page 26: ......

Page 27: ......

Page 28: ...1 008 1048 bd 20220914 06 en...