Preparation methods according to ISO 17664
59
The medical device has a max. temperature resistance of 138℃.
KaVo recommends for example
- STERIclave B 2200/ 2200P by KaVo
- Citomat/ K-series by Getinge
Autoclave with a triple prevacuum for at least 4 minutes at 134°C ± 1
(273 °F ± 1.8)
Autoclave using the gravitation method for at least 10 minutes at 134°C ±
1
(273 °F ± 1.8)
Autoclave using the gravitation method for at least 60 minutes at 121°C ±
1
(250 °F ± 1.8)
Follow the manufacturer's instructions for use.