Preparation methods according to ISO 17664
40
Preparation at the site of use
CAUTION
Hazard from nonsterile products.
An infection hazard exists from contaminated medical devices.
▶
Observe suitable personal protective measures.
▶
Remove residual cement, composite or blood at the site of use.
▶
The medical device must be dry when transporting it to be prepared.
Do not place it in a solution or the like.
▶
The medical device should be prepared as close to the treatment time
as possible.
▶
Remove carbide cutters or diamond burs from the medical device.