Troubleshooting
36
Exchanging the O-rings on the coupling on the supply hose
CAUTION
Hazard from improper care of the O-rings.
Improper care of the O-ring can cause the medical device to partially or
completely malfunction.
▶
Do not use Vaseline or other grease or oil.
Note
The O-rings on the coupling may only be lubricated with a cotton ball wet
with KAVOspray.
▶
Press the O-ring between your fingers to form a loop.
▶
Shove the O-ring to the front, and remove it.
▶
Insert new O-rings into the grooves.