6 Reprocessing steps in accordance with ISO 17664 | 6.6 Sterilisation
18 / 22
6.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / ISO 17665-1
NOTICE
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶
Before each sterilisation cycle, service the medical device with KaVo care
products.
NOTICE
Contact corrosion due to moisture.
Damage to product.
▶
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
The KaVo medical device has a maximum temperature resistance up to 138
℃
(280.4 °F).
Sterilisation parameters:
Select a suitable procedure (depending on the available autoclave) from the fol-
lowing sterilisation processes:
Steriliser with triple pre-vacuum:
▪ at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
Steriliser using the gravity method:
▪ at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4 °F)
▶
Use according to the manufacturer's Instructions for Use.
6.7 Storage
Prepared products must be stored appropriately in a dry, dark, cool room such
that they are protected from germs (as far as possible) and dust.
Note
Comply with the expiry date of the sterilised items.