Instructions for use INTRA Motorhandstück PROPHYwiz 181 P - REF 1.003.2278
6 Reprocessing steps in accordance with ISO 17664 | 6.4 Care products and systems - Servicing
CAUTION
Incorrect servicing of prophylaxis heads.
Infection.
▶
Comply with the manufacturer's instructions concerning the reprocessing of
prophylaxis heads.
Note
KaVo guarantees the proper function of KaVo products only if the care
products listed as accessories are used, since these were tested for proper
use on our products.
6.4.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Take the head off.
▶
Unscrew the motor handpiece from the hose.
▶
Cover the medical device with the KaVo Cleanpac bag.
▶
Use an appropriate cannula and apply spray into the smaller of the two
large tubes.
▶
Press the spray key for 1 to 2 seconds.
Note
Additional servicing is required if the working speed decreases due to con-
tamination or particles being present in the air supply system.
Chuck care
KaVo recommends servicing the chucking system once weekly.
▶
Remove the head and spray into the opening using the tip of the spray
nipple.
▶
Press the spray key for 1 to 2 seconds.
6.4.2 Servicing with KaVo QUATTROcare PLUS
Cleaning and servicing device with expansion pressure for internal cleaning of
inorganic residues and optimum care.
(no validated cleaning of the interior according to German Robert Koch Institute
(RKI) requirements)
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Take the head off.
▶
Unscrew the motor handpiece from the hose.
▶
Service the product in the QUATTROcare PLUS.
15 / 22